FDA Adverse Event Malfunction Summary report: N

HENRY SCHEIN INC

MDR report key: 22292405 · Received June 19, 2025

Report

Report Number
3003444492-2025-00004
Event Type
Malfunction
Date Received
June 19, 2025
Date of Event
May 8, 2025
Report Date
June 19, 2025
Manufacturer
PRIMA DENTAL GROUP
Product Code
EJL
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

COMPLAINANT UNABLE TO PROVIDE WHICH PART NUMBER CAUSED THE INCIDENT, NOR COULD THEY PROVIDE ANY LOT NUMBERS. DEVICES ARE UNABLE TO BE RETURNED TO THE MANUFACTURER FOR INVESTIGATION.

Additional Manufacturer Narrative · 0

COMPLAINANT UNABLE TO PROVIDE WHICH PART NUMBER CAUSED THE INCIDENT, NOR COULD THEY PROVIDE ANY LOT NUMBERS. DEVICES ARE UNABLE TO BE RETURNED TO THE MANUFACTURER FOR INVESTIGATION. INITIAL CHECK THROUGH THE RETURNED BOX FOUND 3 DIFFERENT PRODUCTS, FG557, FG558 AND FG330. DEVICE WERE RETURNED AND AN INVESTIGATION WAS ABLE TO BE CARRIED OUT BURS TESTED ON THE TEST RIG AND NO BURS WERE FOUND TO LEAVE THE HANDPIECE. DIMENSIONAL CHECK WAS PERFORMED ON RETURNED PARTS AND THE RESULTS WERE SIMILAR COMPARED TO APPROVED PARTS FROM STOCK. INVESTIGATION HAS BEEN MADE ON THE MANUFACTURING RECORDS FOR THE INFORMED LOT NUMBER AND NO ABNORMALITY WAS FIND. THEREFORE, THE COMPLAINT IS CONSIDERED UNJUSTIFIED.

Description of Event or Problem · 0

THE CUSTOMER REPORTS THAT ON (B)(6) 2025 A 6 YEAR OLD PATIENT THE DOCTOR WAS DOING CROWNS ON S AND T W/PULP. WHILE THE DOCTOR WAS DRILLING, THE BUR CAME OUT AND CUT THE INSIDE OF THE CHILD'S CHEEK. THE DR. WAS ABLE TO STOP THE BLEEDING, AND WANTED TO SEND THE CHILD TO THE HOSPITAL, BUT AFTER SPEAKING TO MOM, THEY WERE GOING TO MONITOR THE CUT OVER A FEW DAYS. MOM WAS CALLED AND SHE STATED THAT IT WAS HEALING. COMMON BURS USED (NO LOT#'S AVAILABLE) FG557 (ITEM 1014491) AND FG330 (ITEM #1010237). THE HANDPIECE IS HS ITEM CODE: 5700108, MAXIMA ELITE MINI.

Description of Event or Problem · 0

THE CUSTOMER REPORTS THAT ON (B)(6) 2025 A 6-YEAR-OLD PATIENT THE DOCTOR WAS DOING CROWNS ON S AND T W/PULP. WHILE THE DOCTOR WAS DRILLING, THE BUR CAME OUT AND CUT THE INSIDE OF THE CHILD'S CHEEK. THE DR. WAS ABLE TO STOP THE BLEEDING, AND WANTED TO SEND THE CHILD TO THE HOSPITAL, BUT AFTER SPEAKING TO MOM, THEY WERE GOING TO MONITOR THE CUT OVER A FEW DAYS. MOM WAS CALLED AND SHE STATED THAT IT WAS HEALING. COMMON BURS USED (NO LOT#'S AVAILABLE) FG557 (ITEM 1014491) AND FG330 (ITEM #1010237). THE HANDPIECE IS HS ITEM CODE: 5700108, MAXIMA ELITE MINI.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1684229 HENRY SCHEIN INC DENTAL BUR EJL PRIMA DENTAL GROUP NOT KNOW NOT KNOW

Patients

Seq Age Sex Outcome Treatment
1 6 YR Unknown