HENRY SCHEIN INC
Report
- Report Number
- 3003444492-2025-00004
- Event Type
- Malfunction
- Date Received
- June 19, 2025
- Date of Event
- May 8, 2025
- Report Date
- June 19, 2025
- Manufacturer
- PRIMA DENTAL GROUP
- Product Code
- EJL
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
COMPLAINANT UNABLE TO PROVIDE WHICH PART NUMBER CAUSED THE INCIDENT, NOR COULD THEY PROVIDE ANY LOT NUMBERS. DEVICES ARE UNABLE TO BE RETURNED TO THE MANUFACTURER FOR INVESTIGATION.
COMPLAINANT UNABLE TO PROVIDE WHICH PART NUMBER CAUSED THE INCIDENT, NOR COULD THEY PROVIDE ANY LOT NUMBERS. DEVICES ARE UNABLE TO BE RETURNED TO THE MANUFACTURER FOR INVESTIGATION. INITIAL CHECK THROUGH THE RETURNED BOX FOUND 3 DIFFERENT PRODUCTS, FG557, FG558 AND FG330. DEVICE WERE RETURNED AND AN INVESTIGATION WAS ABLE TO BE CARRIED OUT BURS TESTED ON THE TEST RIG AND NO BURS WERE FOUND TO LEAVE THE HANDPIECE. DIMENSIONAL CHECK WAS PERFORMED ON RETURNED PARTS AND THE RESULTS WERE SIMILAR COMPARED TO APPROVED PARTS FROM STOCK. INVESTIGATION HAS BEEN MADE ON THE MANUFACTURING RECORDS FOR THE INFORMED LOT NUMBER AND NO ABNORMALITY WAS FIND. THEREFORE, THE COMPLAINT IS CONSIDERED UNJUSTIFIED.
THE CUSTOMER REPORTS THAT ON (B)(6) 2025 A 6 YEAR OLD PATIENT THE DOCTOR WAS DOING CROWNS ON S AND T W/PULP. WHILE THE DOCTOR WAS DRILLING, THE BUR CAME OUT AND CUT THE INSIDE OF THE CHILD'S CHEEK. THE DR. WAS ABLE TO STOP THE BLEEDING, AND WANTED TO SEND THE CHILD TO THE HOSPITAL, BUT AFTER SPEAKING TO MOM, THEY WERE GOING TO MONITOR THE CUT OVER A FEW DAYS. MOM WAS CALLED AND SHE STATED THAT IT WAS HEALING. COMMON BURS USED (NO LOT#'S AVAILABLE) FG557 (ITEM 1014491) AND FG330 (ITEM #1010237). THE HANDPIECE IS HS ITEM CODE: 5700108, MAXIMA ELITE MINI.
THE CUSTOMER REPORTS THAT ON (B)(6) 2025 A 6-YEAR-OLD PATIENT THE DOCTOR WAS DOING CROWNS ON S AND T W/PULP. WHILE THE DOCTOR WAS DRILLING, THE BUR CAME OUT AND CUT THE INSIDE OF THE CHILD'S CHEEK. THE DR. WAS ABLE TO STOP THE BLEEDING, AND WANTED TO SEND THE CHILD TO THE HOSPITAL, BUT AFTER SPEAKING TO MOM, THEY WERE GOING TO MONITOR THE CUT OVER A FEW DAYS. MOM WAS CALLED AND SHE STATED THAT IT WAS HEALING. COMMON BURS USED (NO LOT#'S AVAILABLE) FG557 (ITEM 1014491) AND FG330 (ITEM #1010237). THE HANDPIECE IS HS ITEM CODE: 5700108, MAXIMA ELITE MINI.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1684229 | HENRY SCHEIN INC | DENTAL BUR | EJL | PRIMA DENTAL GROUP | NOT KNOW | NOT KNOW |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 6 YR | Unknown |