FDA Adverse Event Malfunction Summary report: N

COLLEAGUE

MDR report key: 2228748 · Received August 31, 2011

Report

Report Number
6000001-2011-21594
Event Type
Malfunction
Date Received
August 31, 2011
Date of Event
June 28, 2011
Report Date
August 6, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE CONDITION WAS CONFIRMED BY SERVICE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE CAUSE WAS DETERMINED TO BE THE FAULTY MAIN BATTERIES. TO CORRECT THIS CONDITION, THE MAIN BATTERIES AND BATTERY HARNESS WERE REPLACED. IF ADDITIONAL INFORMATION OR SAMPLES BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING A REVIEW OF THE EVENT HISTORY BY BAXTER PERSONNEL, A COLLEAGUE INFUSION PUMP EXPERIENCED TWO OCCURRENCES OF FAILURE CODE 589:317:1061:0000. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THIS EVENT. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.09.90. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1