FDA Adverse Event Injury Summary report: N

HAMILTON GALILEO

MDR report key: 222847 · Received May 10, 1999

Report

Report Number
MW1016277
Event Type
Injury
Date Received
May 10, 1999
Date of Event
May 7, 1999
Report Date
May 10, 1999
Manufacturer
HAMILTON MEDICAL, INC.
Product Code
CBK
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
FL, US
Reporter Occupation
RESPIRATORY THERAPIST

Narratives

Description of Event or Problem · 1

A MONITOR TECH RESPONDED TO A LOW OXYGEN SATURATION ALARM OF 54%. THE VENTILATOR CIRCUIT WAS FOUND DISCONNECTED FROM THE PT AND THE PT WAS CYANOTIC AND GASPING. THE PT WAS MANUALLY VENTILATED WITH 100% OXYGEN, AND ONCE COLOR IMPROVED WAS PLACED BACK ON THE VENTILATOR. THE VENTILATOR ALARMED DISCONNECT, HOWEVER NURSING PERSONNEL DID NOT SENSE THE URGENCY OF THE VENTILATOR ALARM. THE VENTILATOR BEEPED FIVE TIMES, PAUSED FOR 5 SECS, AND BEEPED AGAIN FIVE TIMES FOLLOWED BY THE PAUSE. OTHER VENTILATORS AT THIS INSTITUTION ALARM A CONTINUOUS TONE WITHOUT INTERRUPTION WHEN THERE IS A DISCONNECT. NURSING HAS BEEN EDUCATED ABOUT THE ALARM SYSTEM, BUT STILL FEEL THE ALARM IS NOT LOUD ENOUGH TO OBTAIN ATTENTION, AND SHOULD BE A CONTINUOUS TONE TO ALERT THE USER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HAMILTON GALILEO VENTILATOR CBK HAMILTON MEDICAL, INC. GALILEO *

Patients

Seq Age Sex Outcome Treatment
1 75 YR Life Threatening