FDA Adverse Event Malfunction Summary report: N

CARESTATION 650

MDR report key: 22279799 · Received June 18, 2025

Report

Report Number
2112667-2025-05030
Event Type
Malfunction
Date Received
June 18, 2025
Date of Event
May 23, 2025
Report Date
June 18, 2025
Manufacturer
BZ
Product Code
BSZ
UDI-DI
00840682103947
PMA / PMN Number
K151570
Removal / Correction Number
SEE H11
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

GE HEALTHCARE (GEHC) REPORTED A FIELD MODIFICATION FOR AN ISSUE THAT PREVENTS EFFECTIVE MECHANICAL VENTILATION IN VOLUME CONTROL VENTILATION (VCV) MODE PER 21 CFR 806 ON 02-APR-2025. THE FDA RECALL NUMBERS ARE: Z-1624-2025, Z-1625-2025, Z-1626-2025, Z-1627-2025, Z-1628-2025, Z-1629-2025, Z-1630-2025, Z-1631-2025, Z-1632-2025, Z-1633-2025, Z-1634-2025, Z-1635-2025, Z-1636-2025, Z-1637-2025, Z-1638-2025. CUSTOMERS WERE SENT A LETTER EXPLAINING THE ISSUE AND PROVIDING INSTRUCTIONS FOR INSPECTING FOR EFFECTIVE VENTILATION IN VOLUME CONTROL VENTILATION (VCV) MODE. GEHC WILL CORRECT ALL DEVICES THAT FAIL THE VENTILATION SCREENING TEST AT NO COST. BLOCK A: NO PATIENT INVOLVEMENT. LEGAL MANUFACTURER: HCS MADISON - 3030 OHMEDA DR, USA MADISON, WI 53718.

Description of Event or Problem · 0

AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION INITIATED BY GE HEALTHCARE (GEHC) ON 02-APR-2025, (RECALL NUMBERS Z-1624-2025, Z-1625-2025, Z-1626-2025, Z-1627-2025, Z-1628-2025, Z-1629-2025, Z-1630-2025, Z-1631-2025, Z-1632-2025, Z-1633-2025, Z-1634-2025, Z-1635-2025, Z-1636-2025, Z-1637-2025, Z-1638-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE THAT PREVENTS EFFECTIVE MECHANICAL VENTILATION IN VOLUME CONTROL VENTILATION (VCV) MODE. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
704131 CARESTATION 650 ANESTHESIA GAS MACHINE BSZ BZ A1 NA 00840682103947

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown