VERSAONE
Report
- Report Number
- 9612501-2025-01550
- Event Type
- Malfunction
- Date Received
- June 18, 2025
- Date of Event
- June 10, 2025
- Report Date
- October 13, 2025
- Manufacturer
- DAVIS & GECK CARIBE LTD
- Product Code
- GCJ
- UDI-DI
- 10884521657687
- PMA / PMN Number
- K130435
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
H3 EVALUATION SUMMARY: MEDTRONIC CONDUCTED AN INVESTIGATION BASED UPON ALL INFORMATION RECEIVED. THE DEVICE WAS AVAILABLE FOR EVALUATION. A VIDEO WAS ALSO PROVIDED. VISUAL INSPECTION NOTED THAT THE OBTURATOR WAS NOT RECEIVED. THE SEAL HOUSING, CIRCULAR SEAL, STOPCOCK, DUCKBILL SEAL AND CANNULA APPEARED INTACT. FUNCTIONAL TESTING NOTED THAT THE DEVICE PASSED THE AIR LEAK TEST. THE DUCKBILL SEAL FAILED DURING MANUAL MANIPULATION USING AN ENDO PEANUT. THE SEAL HOUSING WAS BROKEN OPEN TO REMOVE THE CIRCULAR SEAL TO CHECK FOR SUBASSEMBLY OVERALL HEIGHT. THE MEASUREMENTS WITH CALIPERS WERE 251, 253, 255, 251, 252 AND 255. IT WAS REPORTED THAT THE SEAL WAS DAMAGED AND THERE WERE LEAKS. THE REPORTED ISSUES WERE CONFIRMED. THE MOST LIKELY CAUSE COULD NOT BE ESTABLISHED FROM THE INFORMATION AVAILABLE. THE MANUFACTURING RECORDS FOR EACH DEVICE ARE THOROUGHLY REVIEWED PRIOR TO RELEASE TO ENSURE THAT IT MEETS ALL MEDTRONIC QUALITY SPECIFICATIONS. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
ADDITIONAL INFORMATION: G3, H3, H6. H3. EVALUATION SUMMARY: MEDTRONIC CONDUCTED AN INVESTIGATION BASED UPON ALL INFORMATION RECEIVED. THE INVOLVED DEVICE WAS NOT RETURNED. HOWEVER, A VIDEO WAS PROVIDED. VISUAL INSPECTION NOTED THAT THE VIDEO SHOWS THE SURGEON REMOVING THE SEAL HOUSING AND PRESSING THE CENTER OF THE SEAL. DUE TO THE QUALITY OF THE VIDEO IT IS NOT POSSIBLE TO CONFIRM THE REPORTED CONDITION. IT WAS REPORTED THAT SEAL WAS DAMAGED AND THERE WAS AN AIR LEAKAGE. THE REPORTED ISSUES WERE NOT CONFIRMED. THE MOST LIKELY CAUSE COULD NOT BE ESTABLISHED FROM THE INFORMATION AVAILABLE. THE MANUFACTURING RECORDS FOR EACH DEVICE ARE THOROUGHLY REVIEWED PRIOR TO RELEASE TO ENSURE THAT IT MEETS ALL MEDTRONIC QUALITY SPECIFICATIONS. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
IT WAS REPORTED THAT DURING A BARIATRIC LAPAROSCOPIC PROCEDURE, THE SEAL WAS DAMAGED THAT RESULTED TO A GAS LEAK FROM THE SEAL OF THE BLADELESS TROCAR WHILE THE DEVICE WAS BEING APPLIED DURING THE HEMOSTASIS STEP. THE TROCAR CANNULA WAS REPLACED TO RESOLVE THE ISSUE AND COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 344603 | VERSAONE | LAPAROSCOPE, GENERAL & PLASTIC SURGERY | GCJ | DAVIS & GECK CARIBE LTD | ONB12STF | J4C2940Y | 10884521657687 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Female |