AUTOSOFT XC
Report
- Report Number
- 3003442380-2025-10774
- Event Type
- Malfunction
- Date Received
- June 18, 2025
- Date of Event
- May 29, 2025
- Report Date
- June 24, 2025
- Manufacturer
- UNOMEDICAL DEVICES S.A. DE C.V.
- Product Code
- FPA
- UDI-DI
- 05705244016620
- PMA / PMN Number
- K032854
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS H11: INVESTIGATION SUMMARY THE INFORMATION IN THIS COMPLAINT (B)(4) HAS BEEN EVALUATED. THE BATCH 6010720 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. INVESTIGATION PROCESS OF THE COMPLAINT WAS CARRIED OUT IN ACCORDANCE WITH WORK INSTRUCTION (WI) GUIDANCE FOR VISUAL TESTING FOR COMPLAINTS AREA VERSION 3 AND WI GUIDANCE FOR FUNCTIONAL TESTING FOR COMPLAINTS AREA VERSION 2 FOR THE CODE LEAKAGE FROM TUBING OR THE INTERFACE BETWEEN THE TUBING AND THE CONNECTORS (REFERS TO CONNECTORS IN BOTH ENDS OF THE TUBING) COMPLAINT INVESTIGATION. PHOTO/SAMPLE WAS NOT PROVIDED. IN ORDER TO TEST THE PRODUCT, THE REFERENCE SAMPLES FROM THE LOT HAVE BEEN REQUESTED. REFERENCE SAMPLES TEST RESULTS: ON THE VISUAL INSPECTION ACCORDING TO WI VERSION 3, WAS PERFORMED AND (B)(6) SAMPLES PASSED THE TEST. FUNCTIONAL TEST AIR FLOW ACCORDING TO WI VERSION 2 ON REFERENCE SAMPLES, (B)(6) SAMPLES PASSED THE TEST. FUNCTIONAL TEST AIR LEAK ACCORDING TO WI VERSION 2 ON REFERENCE SAMPLES, 10/10 SAMPLES PASSED THE TEST. A BATCH RECORD REVIEW WAS CONDUCTED RESULTING IN THE FOLLOWING: PACKAGING LOT: THE LOT 6010720 WAS PACKAGING ACCORDING TO THE WI VERSION 120, IN THE LINE INSET 4, ON (B)(6) 2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DEVICE HISTORY RECORD (DHR) SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WERE IDENTIFIED; NO MAINTENANCE EVENTS WERE RECORDED. TRENDING: A QUERY WAS RUN IN DATABASE ON (B)(6) 2025: AGAINST MALFUNCTION CODE LEAKAGE FROM TUBING OR THE INTERFACE BETWEEN THE TUBING AND THE CONNECTORS (REFERS TO CONNECTORS IN BOTH ENDS OF THE TUBING) AND LOT 6010720 AND NO OTHER COMPLAINT HAS BEEN REGISTERED IN DATABASE FOR THE SAME LOT AND MALFUNCTION CODE. CONCLUSION SUMMARY OF THE RELATED EVENT: AS A RESULT OF THE FOLLOWING: ON THE INVESTIGATION ON REFERENCE SAMPLES, NO ISSUE WERE FOUND RELATED TO LEAK IN TUBING, HARM NO REPORTABLE, NO DEFECT ON TESTS, NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION, NO MORE COMPLAINT WAS RECEIVED ON THE LOT IN QUESTION AND MALFUNCTION, NO FURTHER ACTION ARE REQUIRED, THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT PATIENT FACED AN INFUSION SET LEAKAGE EVENT ON (B)(6) 2025 AT T-LOCK. INFUSION SET WAS USED FOR 30 HOURS. PATIENT REPLACED INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY NO FURTHER INFORMATION AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1527380 | AUTOSOFT XC | UNO INSET I 60/6 GREY TCAP 10PK INT | FPA | UNOMEDICAL DEVICES S.A. DE C.V. | 1001680 | 6010720 | 05705244016620 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 28 YR | Male |