ACCESS 2 IMMUNOASSAY SYSTEM
Report
- Report Number
- 2122870-2011-02874
- Event Type
- Malfunction
- Date Received
- August 12, 2011
- Date of Event
- July 25, 2008
- Report Date
- July 28, 2008
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K922823/A007
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
SVC WAS NOT DISPATCHED AS THE CUSTOMER DID NOT QUESTION SYSTEM PERFORMANCE. TROPONIN I (ACCUTNI) QUALITY CONTROL (QC) LEVELS ONE AND THREE RESULTED WITHIN SPECIFICATIONS ON THE DAYS THE PT SAMPLES WERE ANALYZED. TROPONIN I QC LEVEL ONE RESULTED OUT OF SPECIFICATIONS (LOW) ON (B)(6) 2008 AND RESULTED WITHIN RANGE UPON RETEST. A SYSTEM CHECK PERFORMED ON (B)(6) 2008 CONFORMED TO SPECIFICATIONS. TESTING AT BECKMAN COULTER CUSTOMER PRODUCT LINE SUPPORT (CPLS) LAB CONFIRMED THE EXISTENCE OF A PT SOURCE INTERFERENT FOR THE SAMPLES RECEIVED FROM THE CUSTOMER. THE INTERFERENCE LIKELY RELATES TO ALKALINE PHOSPHATASE. THIS INTERFERING COMPOUND CAUSES REPRODUCIBLE FALSE POSITIVE RESULTS, BUT IS DISTINCT FROM HETEROPHILE ANTIBODIES. FOR ASSAYS EMPLOYING MOUSE ANTIBODIES, THE POSSIBILITY EXISTS FOR INTERFERENCE BY HUMAN ANTI-MOUSE ANTIBODIES (HAMA) IN THE SAMPLE. HUMAN ANTI-MOUSE ANTIBODIES MAY BE PRESENT IN SAMPLES FROM PTS WHO HAVE RECEIVED IMMUNOTHERAPY OR DIAGNOSTIC PROCEDURES UTILIZING MONOCLONAL ANTIBODIES OR IN INDIVIDUALS WHO HAVE BEEN REGULARLY EXPOSED TO ANIMALS. ADDITIONALLY, OTHER HETEROPHILE ANTIBODIES, SUCH AS HUMAN ANTI-GOAT ANTIBODIES, MAY BE PRESENT IN PT SAMPLES. THE ACCESS ACCUTNI RESULTS SHOULD BE INTERPRETED IN LIGHT OF THE TOTAL CLINICAL PRESENTATION OF THE PT, INCLUDING: SYMPTOMS, CLINICAL HISTORY, CLINICAL EXAMINATION, ELECTROCARDIOGRAM (ECG), DATA FROM ADDITIONAL TESTS, AND OTHER APPROPRIATE INFO. MEDICAL DECISIONS SHOULD NOT BE BASED ON A SINGLE ACCUTNI DETERMINATION AT ONE TIME POINT. THIS REPORTABLE EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW OF PRODUCT COMPLAINTS CONDUCTED FROM (B)(4) 2008 THROUGH (B)(4) 2010 FOR ADDITIONAL REPORTABLE EVENTS. THIS MEDWATCH REPORT IS RELATED TO MDR 2122870-2011-02820.
THE AFFILIATE ALLEGED THE CUSTOMER REPORTED ELEVATED TROPONIN I (ACCUTNI) RESULTS, ABOVE THE ACUTE MYOCARDIAL INFARCTION (AMI) CUTOFF, FOR ONE PT INVOLVING ACCESS 2 IMMUNOASSAY SYSTEM. THIS IS REPORT NUMBER ONE OF TWO REFERENCING SYSTEM SERIAL NUMBER (B)(4). THE ELEVATED RESULTS WERE REPRODUCIBLE AND DID NOT CORRELATE WITH THE PT'S CLINICAL CONDITION. THE PT EXHIBITED NO CLINICAL SIGNS OF HEART FAILURE. THE ELEVATED RESULTS WERE RELEASED OUT OF THE LAB. THERE WAS NO REPORT OF PT INJURY OR CHANGE IN PT TREATMENT ASSOCIATED WITH THIS EVENT. THE CUSTOMER SUPPLIED BECKMAN COULTER, INC. IN (B)(4) WITH THE PT SAMPLES FOR FURTHER ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACCESS 2 IMMUNOASSAY SYSTEM | JJE | BECKMAN COULTER, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | ACCESS ACCUTNI |