FDA Adverse Event Malfunction Summary report: N

ENDOWRIST

MDR report key: 22243564 · Received June 17, 2025

Report

Report Number
2955842-2025-25616
Event Type
Malfunction
Date Received
June 17, 2025
Date of Event
April 11, 2025
Report Date
June 5, 2025
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874112311
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) HAS RECEIVED THE PERMANENT CAUTERY HOOK INSTRUMENT ASSOCIATED WITH THIS COMPLAINT AND COMPLETED INVESTIGATIONS. THE INSTRUMENT WAS FOUND TO HAVE A BROKEN CONDUCTOR WIRE AT THE WELD ON YAW PULLEY. THE INSTRUMENT FAILED THE ELECTRICAL CONTINUITY TEST. NO SIGNS OF THERMAL DAMAGE WERE OBSERVED. A REVIEW OF THE SUBMITTED IMAGE WAS PERFORMED, AND ADDITIONAL INFORMATION WAS RECEIVED: THE CUSTOMER PROVIDED AN IMAGE OF THE INSTRUMENT; HOWEVER, THE POTENTIAL ISSUE / DAMAGE COULD NOT BE SEEN NOR DETERMINED BASED ON THE IMAGE.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED ESOPHAGECTOMY SURGICAL PROCEDURE, THE PERMANENT CAUTERY HOOK INSTRUMENT DID NOT CAUTERIZE/INSUFFICIENT ENERGY. THERE WERE DIFFICULTIES WITHIN WRIST/ENERGY USING THE INSTRUMENT. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE INSTRUMENT WAS NOT WORKING WHEN THE SURGEON WAS ABOUT TO USE IT DURING THE PROCEDURE. THEY DID NOT OBSERVE ANY PHYSICAL DAMAGE OR BROKEN CABLES, SO IT COULD BE RELATED MORE WITH ENERGY, HOWEVER IT COULD ALSO BE A MISTAKE DURING THE INSTRUMENT PREPARATION AND THAT IT COULD HAVE NO DAMAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1630121 ENDOWRIST PERMANENT CAUTERY HOOK NAY INTUITIVE SURGICAL, INC 470183-18 K17241212 0101 00886874112311

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown DA VINCI INSTRUMENTS AND ACCESSORIES