FDA Adverse Event Injury Summary report: N

RESTORE ULTRA

MDR report key: 2223857 · Received August 18, 2011

Report

Report Number
3004209178-2011-06535
Event Type
Injury
Date Received
August 18, 2011
Date of Event
January 1, 2011
Report Date
October 4, 2018
Manufacturer
MDT PUERTO RICO OPERATIONS CO., JUNCOS
Product Code
LGW
PMA / PMN Number
P840001
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
NOT APPLICABLE
Health Professional
N

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT, AFTER THE PT'S IMPLANTABLE NEUROSTIMULATOR WAS IMPLANTED IN SURGERY, IMPEDANCE MEASUREMENTS OF ELECTRODES 0/13 AND 0/14 WERE OUT OF RANGE. THE LEAD THAT HAD JUST BEEN IMPLANTED DURING THE SAME PROCEDURE (MODEL UNK, LOT# UNK) WAS REMOVED AND REPLACED. AFTER THE NEW LEAD (MODEL 39565, LOT# V626185009) WAS PLACED, ALL IMPEDANCE READINGS WITH REFERENCE ELECTRODE 0 WERE GREATER THAN 10,000 OHMS. THE REFERENCE ELECTRODE WAS CHANGED AND THE IMPEDANCE MEASUREMENTS WERE WITHIN NORMAL LIMITS (BETWEEN 700 AND 1,200 OHMS). THE MFR REP PROGRAMMED THE PT'S DEVICE. THE PT FELT STIMULATION AND WAS "SATISFIED". THE ISSUE WAS RESOLVED. SEE MFR REPORT # 3004209178-2011-06533 FOR INFO RELATED TO THE PREVIOUSLY IMPLANTED NEUROSTIMULATOR AND ASSOCIATED EVENTS DURING THE SAME SURGICAL PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RESTORE ULTRA LGW MDT PUERTO RICO OPERATIONS CO., JUNCOS 37712 NA

Patients

Seq Age Sex Outcome Treatment
1 48 YR Female Required Intervention EXPLANTED.| EXPLANTED: | EXPLANTED: | IMPLANTED.| IMPLANTED: | IMPLANTED: | LEAD: MODEL 39565, LOT# V626185009.| LEAD: MODEL 39565-651, LOT # V582460011.| LEAD: MODEL UNK, LOT# UNK.| PROGRAMMER: MODEL 37743, LOT# NKE143159N.| RECHARGER: MODEL 37752, LOT# NKA137455N.| EXPLANTED:| RECHARGER: MODEL 37752, LOT# NKA137455N| LEAD: MODEL 39565, LOT# V626185009| IMPLANTED:| PROGRAMMER: MODEL 37743, LOT# NKE143159N| EXPLANTED:| LEAD: MODEL UNK, LOT# UNK| IMPLANTED: