FDA Adverse Event Malfunction Summary report: N

AU643-02E CLINICAL CHEMISTRY ANALYZER

MDR report key: 2223810 · Received August 26, 2011

Report

Report Number
2050012-2011-04752
Event Type
Malfunction
Date Received
August 26, 2011
Date of Event
July 26, 2011
Report Date
July 26, 2011
Manufacturer
BECKMAN COULTER MISHIMA K.K.
Product Code
JJE
PMA / PMN Number
K961274
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A BEC FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED FOR THIS EVENT AND FOUND THE AU640 ANALYZER TO BE OPERATIONAL AND PERFORMING TO SPECIFICATIONS. THE SOURCE OF THE BURNING SMELL WAS SOLELY FROM THE UNINTERRUPTIBLE POWER SUPPLY (UPS). THE FSE DISCONNECTED THE SYSTEM FROM THE UPS UNTIL A REPLACEMENT WAS PROVIDED. (B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) REPORTING A BURNING SMELL COMING FROM THE (B)(4) CLINICAL CHEMISTRY ANALYZER. NO INJURY WAS REPORTED IN THIS EVENT AS THE CUSTOMER MERELY NOTICED A SMOKE ODOR FROM THE SYSTEM AND POWERED IT OFF.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AU643-02E CLINICAL CHEMISTRY ANALYZER CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER MISHIMA K.K. AU643-02E N/A

Patients

Seq Age Sex Outcome Treatment
1