FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2223763 · Received August 25, 2011

Report

Report Number
6000001-2011-20720
Event Type
Malfunction
Date Received
August 25, 2011
Date of Event
July 31, 2011
Report Date
August 15, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915522
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PR
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A REQUEST FOR THE RETURN OF THE DEVICE HAS BEEN MADE. SHOULD THE DEVICE BE RECEIVED BY BAXTER FOR EVALUATION, A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF AN EVALUATION OR IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A FLOGARD INFUSION PUMP WITH A BROKEN DOOR WAS CONFIRMED DURING PRODUCT EVALUATION. THIS CONDITION WAS CAUSED BY A BROKEN DOOR LATCH. THE DOOR LATCH, PUMP HEAD DOOR AND THE OCCLUSION DETECTORS WERE REPLACED TO CORRECT THE REPORTED CONDITION. BAXTER DISCONTINUED SERVICE AND CEASED ALL DESIGN RELATED SUPPORT ON FLO-GARD (B)(4) IN THE U.S. REGION AS OF (B)(6) 2010. BAXTER WILL CONTINUE TO COLLECT PRODUCT COMPLAINTS, BUT PERIODIC MONITORING/TRENDING AND ANALYSES OF THE U.S. COMPLAINTS FOR PRODUCT IMPROVEMENTS WILL NO LONGER BE REQUIRED. BAXTER WILL CONTINUE TO MONITOR AND REPORT ON DEATH AND SERIOUS INJURY EVENTS AND FOLLOW EXISTING ESCALATION PROCESSES TO ASSESS THE IMPACT ON PATIENT SAFETY.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE CONTACTED A TECHNICAL SERVICE REPRESENTATIVE TO REPORT A FLOGARD INFUSION PUMP WITH A BROKEN LATCH DOOR; THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. IT IS UNKNOWN WHEN THIS EVENT OCCURRED OR IF THERE WAS PATIENT INVOLVEMENT. THERE WAS NO REPORTED PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE EVENT IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1