FDA Adverse Event Malfunction Summary report: N

COLLEAGUE

MDR report key: 2223761 · Received August 25, 2011

Report

Report Number
6000001-2011-20716
Event Type
Malfunction
Date Received
August 25, 2011
Date of Event
August 1, 2011
Report Date
August 16, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. A REQUEST FOR THE RETURN OF THE DEVICE HAS BEEN MADE. SHOULD THE DEVICE BE RECEIVED BY BAXTER FOR EVALUATION, A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF AN EVALUATION OR IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP THAT FAILED WAS CONFIRMED DURING PRODUCT EVALUATION. THE QUALITY ENGINEER HAS DETERMINED THAT THIS CONDITION WAS A RESULT OF A 810:11 FAILURE CODE AND DETERMINED THE CAUSE TO BE NOT IDENTIFIED. SINCE NO ASSIGNABLE CAUSE WAS PROVIDED DURING PRODUCT EVALUATION, NO REPAIR WAS NECESSARY FOR THE REPORTED CONDITION. ADDITIONAL INFORMATION: A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. A DEVICE HISTORY REVIEW WAS PERFORMED AND NO ABNORMALITY WAS OBSERVED IN THE MANUFACTURING OF THIS DEVICE. THIS IS INVOLVING A PUMP WITH SOFTWARE VERSION 5.09.90 WHICH IS CATEGORIZED AS A COLLEAGUE 2006.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE CONTACTED A TECHNICAL SERVICE REPRESENTATIVE TO REPORT A COLLEAGUE INFUSION PUMP THAT FAILED; THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. IT IS UNKNOWN WHEN THIS EVENT OCCURRED. IT IS UNKNOWN IF THERE WAS PATIENT INVOLVEMENT. THERE WAS NO REPORTED PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE EVENT IN ASSOCIATION WITH THIS EVENT. THE SOFTWARE VERSION IS CURRENTLY UNKNOWN AT THIS TIME. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1