FREESTYLE LITE
Report
- Report Number
- 2954323-2011-04295
- Event Type
- Malfunction
- Date Received
- August 25, 2011
- Date of Event
- July 26, 2011
- Report Date
- August 25, 2011
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
RETAIN SAMPLES FROM TEST STRIP LOT 0822524 HAVE BEEN TESTED AND ONE OF EIGHT RETAIN VIALS WAS CONFIRMED TO ALSO PRODUCE LOW RESULTS WITH CONTROL SOLUTION TESTING. FURTHER OBSERVATION OBSERVED A SCRATCH IN THE CHANNEL OF THE CARBON SIDE OF THE STRIP. THE SCRATCH COMPLETELY SEVERS THE CHANNEL CAUSING AN ELECTRICAL "OPEN". THE LOCATION OF THE SCRATCH LEAVES HALF OF THE CARBON AREA TO CONDUCT CHARGE DURING THE REACTION, POTENTIALLY RESULTING IN A LOW READING. THESE ERRONEOUS RESULTS MAY OCCUR IN THE "C" ZONE OF THE PARKES ERROR GRID, AND AS SUCH, HAVE THE POTENTIAL TO BE CLINICALLY SIGNIFICANT. REMEDIAL ACTION 2954323-06/10/09-003-C WAS REPORTED TO THE (B)(6) DISTRICT OFFICE ON 10 JUN 2009. IT SHOULD BE NOTED THAT THE REPORTED TEST STRIPS EXPIRED IN AUGUST 2010.
CUSTOMER REPORTED RECEIVING A READING LOWER THAN HOW THEY FELT ON THEIR ADC METER WHILE USING FREESTYLE LITE BLOOD GLUCOSE TEST STRIPS. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FREESTYLE LITE | GLUCOSE TEST STRIPS | NBW | 0822524 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |