PRISM HTLV-I/HTLV-II
Report
- Report Number
- 1415939-2011-00541
- Event Type
- Malfunction
- Date Received
- August 25, 2011
- Date of Event
- August 8, 2011
- Report Date
- August 10, 2011
- Manufacturer
- ABBOTT LABORATORIES
- Product Code
- MTP
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
AN INVESTIGATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE. (B)(4).
THE CUSTOMER RETURNED FIVE PATIENT SAMPLES TO BE USED IN THIS EVALUATION. WESTERN BLOT TESTING WAS PERFORMED ON THESE SAMPLES AND FOUR OF THE SAMPLES GENERATED NEGATIVE RESULTS WHILE ONE SAMPLE GENERATED INDETERMINATE RESULTS (PRESENCE OF P19 BAND). THE WESTERN BLOT RESULT FOR THIS DISCREPANT SAMPLE DOES NOT MEET THE PRISM PACKAGE INSERT CRITERIA FOR INTERPRETATION AS POSITIVE FOR ANTI-HTLV. AN EVALUATION OF FIELD DATA REPORTED TO THE PRISM METRICS DATABASE FOR PRISM HTLV-I/HTLV-II REAGENT LOT 01876M500 WAS NEXT PERFORMED. (B)(4). THE PRISM HTLV-I/HTLV-II PACKAGE INSERT (COMMODITY NUMBER 34-6646/R04) CONTAINS INFORMATION TO ADDRESS THE CUSTOMER'S CURRENT ISSUE. BASED ON THE RESULTS OF THE CURRENT EVALUATION, THE PRISM HTLV-I/HTLV-II REAGENT KIT LOT 01876M500 IS PERFORMING ACCORDING TO PRODUCT LABEL CLAIMS. A PRODUCT DEFICIENCY WAS NOT IDENTIFIED.
THE CUSTOMER STATES THAT A SAMPLE WAS RECEIVED FOR PLATELETPHERESIS FROM ONE DONOR AND GENERATED FALSE (B)(6) RESULTS WITH THE PRISM HTLV REAGENT LOT 01876M500 (CUT-OFF VALUE IS 1.00 S/CO): SID: (B)(6) PRISM 01143: LOT 01876M500, (B)(6) PRISM 01401: LOT 01876M500, (B)(6) PRISM 0114: LOT 01920M500, (B)(6) . THE ISSUE OCCURRED ON TWO DIFFERENT PRISM ANALYZERS WITH THE SAME REAGENT LOT. WHEN THE CUSTOMER SWITCHED TO REAGENT LOT 01920M500, THE EXPECTED (B)(6) RESULTS WERE GENERATED. THERE IS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PRISM HTLV-I/HTLV-II | MTP | ABBOTT LABORATORIES | 01876M500 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | PRISM ANALYZER LN:06A36-04 SN (B)(4)| PRISM ANALYZER LN:06A36-04 SN (B)(4)| PRISM ANALYZER LN:06A36-04 SN (B)(4)| PRISM HTLVI/HTLVII LN:06E50-68 LOT: 01920M500| PRISMHTLV-I/IILN:06E50-68LOT: 01920M500| PRISM ANALYZER LN:06A36-04 SN (B)(4) |