FDA Adverse Event Malfunction Summary report: N

COLLEAGUE

MDR report key: 2223606 · Received August 25, 2011

Report

Report Number
6000001-2011-20678
Event Type
Malfunction
Date Received
August 25, 2011
Date of Event
July 31, 2011
Report Date
August 19, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).A REQUEST FOR THE RETURN OF THE DEVICE HAS BEEN MADE. SHOULD THE DEVICE BE RECEIVED BY BAXTER FOR EVALUATION, A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF AN EVALUATION OR IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH FAILURE CODE 814:04 WAS CONFIRMED DURING PRODUCT EVALUATION. THE ROOT CAUSE OF THE REPORTED PROBLEM WAS DETERMINED TO BE A FAULTY AIL (AIR IN LINE) BOARD. APPROPRIATE ACTION WAS TAKEN TO CORRECT THE REPORTED PROBLEM BY REPLACING THE AIL PCB (PRINTED CIRCUIT BOARD). BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED. THIS IS INVOLVING A PUMP WITH SOFTWARE VERSION 5.09.90 WHICH IS CATEGORIZED AS A COLLEAGUE 2006.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED A COLLEAGUE INFUSION PUMP WITH A FAILURE CODE 814:04. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. IT IS UNKNOWN WHEN THE EVENT OCCURRED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT, INJURY, OR MEDICAL INTERVENTION NECESSARY. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE SOFTWARE VERSION OF THIS PUMP IS CURRENTLY UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1