FDA Adverse Event Malfunction Summary report: N

PHYSIO COUCH

MDR report key: 2223532 · Received July 21, 2011

Report

Report Number
3003984900-2011-00023
Event Type
Malfunction
Date Received
July 21, 2011
Date of Event
May 26, 2011
Report Date
June 21, 2011
Manufacturer
ARJOHUNTLEIGH
Product Code
FRK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EI
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BEING FILED UNDER EXEMPTION (B)(6) BY (B)(6) ON BEHALF OF THE MFR ARJOHUNTLEIGH, A BRANCH OF (B)(4). PLEASE NOTE THAT PREVIOUS MEDWATCH REPORTS FOR THIS PRODUCT MAY HAVE BEEN SUBMITTED FOR THE MFG SITE HUNTLEIGH HEALTHCARE LTD (UNDER REGISTRATION #(B)(4)). AS OF 2011, THAT NUMBER WAS DE-ACTIVATED DUE TO THE SITE NO LONGER SHIPPING PRODUCT TO THE USA. GOING FORWARD, COMPLAINTS RELATED TO THIS PRODUCT ARE TO BE HANDLED BY ARJOHUNTLEIGH, A BRANCH OF (B)(4) AND ANY MEDWATCH REPORTS WILL BE SUBMITTED UNDER REGISTRATION #(B)(4). THE FIXINGS HOLDING THE BACK REST IN PLACE ARE NOT THE ORIGINAL PARTS OF THE COUCH AND ARE NOT APPROVED AKRON PARTS. THE COUCH HAS BEEN MAINTAINED BY THIRD PARTY - WHO HAS NOT ADHERED TO AVAILABLE IFU AND USED INCORRECT PARTS. THE HEAD SECTION WAS BOLTED TO THE TOP FRAME BY THE CLIENT, INSTEAD OF USING THE CORRECT 10MM HINGE PINS, WHICH ARE READILY AVAILABLE. WHEN ARJOHUNTLEIGH SUPERVISOR ATTENDED THE SITE TO CARRY OUT THE INVESTIGATION, IT WAS FOUND THAT THE COUCH HAD BEEN REPAIRED AGAIN BY A THIRD PARTY, BUT NOT USING THE CORRECT AKRON PARTS; THE REPLACEMENT PARTS ARE NOT SUFFICIENT TO HOLD THE BACKREST. THE CORRECT PARTS HAVE NOW BEEN ORDERED BY ARJOHUNTLEIGH. THIRD PARTY COMPANIES SERVICE THE DEVICES, AND WHILE THE SERVICE MANUALS ARE READILY AVAILABLE, THE MFR HAS NO CONTROL OVER THE PARTS USED. THERE HAVE BEEN NO SIMILAR COMPLAINTS REPORTED WHERE THERE WERE UNAPPROVED COMPONENTS USED BY THIRD PARTY SERVICE COMPANIES LEADING TO THE REPORTED INCIDENT WITHIN THE PAST 15 MONTHS. DESPITE THIS, THE MFR WILL CONTINUE TO MONITOR FUTURE COMPLAINTS IN ORDER TO IDENTIFY ANY TRENDS.

Description of Event or Problem · 1

WHEN A LARGE PT SAT ON THE COUCH AND LEANED BACK ONTO THE BACK REST, BOTH BOLT FIXINGS ON THE BACK OF THE BACKREST SHEARED AND THE PT FELL BACKWARDS. THERE WERE NO INJURIES SUSTAINED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PHYSIO COUCH CLINICAL COUCH FRK ARJOHUNTLEIGH 3222

Patients

Seq Age Sex Outcome Treatment
1 Other