FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2223526 · Received August 25, 2011

Report

Report Number
1423500-2011-11249
Event Type
Malfunction
Date Received
August 25, 2011
Date of Event
August 4, 2011
Report Date
August 4, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS COMPLAINT FOR A LOW DRAIN VOLUME ALARM WAS CONFIRMED AND THE ROOT CAUSE WAS DETERMINED TO BE AIR IN THE PATIENT LINE. THE SAMPLE WAS NOT RETURNED TO BAXTER, THEREFORE NO EVALUATION OR BATCH REVIEW COULD BE PERFORMED. THIS ISSUE IS BEING INVESTIGATED THROUGH CAPA.

Additional Manufacturer Narrative · 1

(B)(4). SAMPLE AVAILABILITY IS UNKNOWN. THE LOT NUMBER IS UNKNOWN; THEREFORE A BATCH REVIEW CANNOT BE PERFORMED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF BAXTER'S INVESTIGATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BAXTER TECHNICAL SERVICES(BTS) REGARDING ASSISTANCE WITH A LOW DRAIN VOLUME ALARM DURING THE INITIAL DRAIN ON THE HOMECHOICE MACHINE(HC). THE HP STATED THERE ARE SEVERAL AIR POCKETS THROUGHOUT THE PATIENT LINE. THE TECHNICAL SERVICE REPRESENTATIVE(TSR) ASSISTED THE HOME PATIENT(HP) TO END THERAPY. THERE WAS PATIENT INVOLVEMENT; HOWEVER, THERE WAS NO INJURY OR MEDICAL INTERVENTION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 60 YR HOMECHOICE