FDA Adverse Event Malfunction Summary report: N

MAQUET OPERATING TABLE SYSTEM TRANSPORTER

MDR report key: 2223499 · Received June 14, 2011

Report

Report Number
8010652-2011-00002
Event Type
Malfunction
Date Received
June 14, 2011
Date of Event
June 9, 2011
Report Date
June 17, 2011
Manufacturer
MAQUET GMBH & CO. KG
Product Code
FPO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

MAQUET SERVICE HAS REPLACED THE BROKEN SCREW AND RETURNED THE TROLLEY TO SERVICE. PRESENTLY, THE ROOT CAUSE FOR THIS INCIDENT IS UNK. MAQUET HAS REQUESTED THE BROKEN SCREW TO BE RETURNED FOR FURTHER ANALYSIS. MAQUET HAS BEEN ABLE TO SIMULATE THE BREAKAGE OF THE RETAINING SCREW REPORTED (WITH THE TROLLEY SUPPORTING A TABLETOP WITH A LOAD OF 180KG) BY REMOVING THE SCREW. HOWEVER, IN TEST, THE TABLETOP NEVER FELL TO THE FLOOR. BECAUSE THE CARRIER FRAME IS FIXED ON THREE POINTS: A LEFT AND RIGHT ROTATION JOINT FOR TRENDELENBURG (ONLY ONE OF THEM WAS BROKEN) AND THE TRENDELENBURG /REVERSE TRENDELENBURG CYLINDER. EVEN IF ONE SCREW FAILED, THE REMAINING JOINTS WOULD HAVE BEEN SUFFICIENT TO SUPPORT THE TABLETOP. GIVEN WHAT IS KNOWN, MAQUET SUSPECTS CUSTOMER MISUSE AS THE ROOT CAUSE OF THE ISSUE. NO SIMILAR INCIDENT WITH THIS TRANSPORTER HAS BEEN REPORTED TO MAQUET SINCE THE LAUNCH OF THIS PRODUCT. (B)(4). SHOULD ADD'L DETAILS BECOME AVAILABLE, A F/U MDR WILL BE SUBMITTED. MAQUET INC. SUBMITS THIS REPORT ON BEHALF OF THE DEVICE MFG FACILITY. MAQUET (B)(4) PROVIDES PRODUCT FAILURE INVESTIGATION, ANALYSIS AND RESOLUTION FOR THE DEVICE DESCRIBED IN THIS REPORT.

Description of Event or Problem · 1

ACCORDING TO THE CUSTOMER, THE FIXATION SCREW BROKE DURING TRANSFER OF THE PT FROM THE TABLE TOP TO HOSPITAL BED. THE TABLE TOP FELL DOWN ON FOOT SIDE. NO INJURY REPORTED. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MAQUET OPERATING TABLE SYSTEM TRANSPORTER STRETCHER, WHEELED; CODE: FPO FPO MAQUET GMBH & CO. KG 1140.62C0

Patients

Seq Age Sex Outcome Treatment
1