SERVO-I
Report
- Report Number
- 8010042-2011-00116
- Event Type
- Malfunction
- Date Received
- June 17, 2011
- Date of Event
- January 28, 2010
- Report Date
- May 20, 2011
- Manufacturer
- MAQUET CRITICAL CARE AB
- Product Code
- CBK
- PMA / PMN Number
- K062793
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE RETURNED AIR GAS MODULE WHICH REGULATES THE INSPIRATORY AIR GAS FLOW TO THE PT WAS INVESTIGATED. THE REPORTED HIGH OXYGEN CONCENTRATION ALARM DURING VENTILATION WAS NOT REPRODUCED. THE TESTS PERFORMED ON THE AIR GAS MODULE REVEALED THAT THE MEMBRANE IN THE NOZZLE UNIT MOUNTED IN THE AIR GAS MODULE WAS STICKY WHICH WAS THE MOST PROBABLE CAUSE OF THE REPORTED FAILURE. THE NOZZLE UNIT, WHICH IS PART OF THE GAS MODULE AND CONSISTS OF A MEMBRANE, A MOUTHPIECE AND A FEATHER SPRING, IS USED FOR REGULATION OF THE GAS FLOW THROUGH THE GAS MODULE. THE MEMBRANE IN THE NOZZLE UNIT IS PRESSED AGAINST A MOUTHPIECE WITH A FEATHER SPRING TO REGULATE THE GAS FLOW THROUGH THE GAS MODULE. THE MEMBRANE STICKINESS MAY CAUSE THE FEATHER SPRING TO GET STUCK TO THE MEMBRANE WHILE BEING PRESSED ONTO IT AND THUS CAUSING A DELAY ON RELEASE. THE DELAY AFFECTS THE REGULATION OF THE GAS FLOW THROUGH THE GAS MODULE AND CAUSES FAILURE DURING PRE-USE CHECK AND, DURING VENTILATION, CAUSES THE DELIVERED GAS FLOW AND PRESSURE TO BE HIGHER THAN SET. A FEATHER SPRING WITH A COARSE SURFACE WAS INTRODUCED IN (B)(6) 2010 TO PREVENT IT FROM GETTING STUCK TO THE MEMBRANE. THE CAUSE TO THE STICKINESS IS WEAR OF THE MEMBRANE IN NOZZLE UNIT. (B)(4).
IT WAS REPORTED THAT THE VENTILATOR ALARMED FOR HIGH OXYGEN CONCENTRATION WHILE IT WAS CONNECTED TO A PT. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SERVO-I | CBK | MAQUET CRITICAL CARE AB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |