SEE H-10
Report
- Report Number
- 2023826-2011-00729
- Event Type
- Malfunction
- Date Received
- August 25, 2011
- Date of Event
- July 26, 2011
- Report Date
- July 27, 2011
- Manufacturer
- STAAR SURGICAL COMPANY
- Product Code
- HQL
- PMA / PMN Number
- P900048
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
RESULTS: VISUAL INSPECTION OF THE RETURNED LENS SHOWED HALF OF THE LENS WAS TORN OFF AND MISSING AND THERE WAS A CLEAR SURGICAL RESIDUE ON THE LENS. (B)(4).
GENDER: UNK. BRAND NAME: SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULAR LENS (ELASTIMIDE). (B)(4). EVALUATION: CONCLUSION: AN INVESTIGATION OF THE ROOT CAUSES OF LENS TEARS, SIMILAR TO THAT STATED IN THIS CLAIM, WERE ADDRESSED IN A CAPA OPENED IN JUNE 2005 (WHICH WAS SUBSEQUENTLY CLOSED). THE CAPA ADDRESSED DELIVERY SYSTEM ISSUES INCLUDING ALL STAGES OF MANUFACTURING OF THE INJECTORS AND CARTRIDGES. PROCESSES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. THE CAPA ALSO ADDRESSED HANDLING ERRORS. ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS. BASED ON THE COMPLAINT HISTORY AND THE EVALUATION OF THE RETURNED PRODUCT, POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND HANDLING ERRORS BY THE CUSTOMER. (B)(4).
THE REPORTER STATED THE AQ2010V THREE PIECE SILICONE LENS SPLIT AS IT WAS INSERTED INTO THE EYE. THE LENS WAS REMOVED WITHOUT ENLARGING THE INCISION AND THERE WAS NO PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SEE H-10 | INTRAOCULAR LENS | HQL | STAAR SURGICAL COMPANY | AQ2010V | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 49 YR | AQ CARTRIDGE-FP: MODEL, LOT NUMBER UNKNOWN| INJECTOR: MODEL MSI-PM, LOT NUMBER UNKNOWN |