FDA Adverse Event Malfunction Summary report: N

SEE H-10

MDR report key: 2223414 · Received August 25, 2011

Report

Report Number
2023826-2011-00729
Event Type
Malfunction
Date Received
August 25, 2011
Date of Event
July 26, 2011
Report Date
July 27, 2011
Manufacturer
STAAR SURGICAL COMPANY
Product Code
HQL
PMA / PMN Number
P900048
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

RESULTS: VISUAL INSPECTION OF THE RETURNED LENS SHOWED HALF OF THE LENS WAS TORN OFF AND MISSING AND THERE WAS A CLEAR SURGICAL RESIDUE ON THE LENS. (B)(4).

Additional Manufacturer Narrative · 1

GENDER: UNK. BRAND NAME: SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULAR LENS (ELASTIMIDE). (B)(4). EVALUATION: CONCLUSION: AN INVESTIGATION OF THE ROOT CAUSES OF LENS TEARS, SIMILAR TO THAT STATED IN THIS CLAIM, WERE ADDRESSED IN A CAPA OPENED IN JUNE 2005 (WHICH WAS SUBSEQUENTLY CLOSED). THE CAPA ADDRESSED DELIVERY SYSTEM ISSUES INCLUDING ALL STAGES OF MANUFACTURING OF THE INJECTORS AND CARTRIDGES. PROCESSES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. THE CAPA ALSO ADDRESSED HANDLING ERRORS. ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS. BASED ON THE COMPLAINT HISTORY AND THE EVALUATION OF THE RETURNED PRODUCT, POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND HANDLING ERRORS BY THE CUSTOMER. (B)(4).

Description of Event or Problem · 1

THE REPORTER STATED THE AQ2010V THREE PIECE SILICONE LENS SPLIT AS IT WAS INSERTED INTO THE EYE. THE LENS WAS REMOVED WITHOUT ENLARGING THE INCISION AND THERE WAS NO PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SEE H-10 INTRAOCULAR LENS HQL STAAR SURGICAL COMPANY AQ2010V N/A

Patients

Seq Age Sex Outcome Treatment
1 49 YR AQ CARTRIDGE-FP: MODEL, LOT NUMBER UNKNOWN| INJECTOR: MODEL MSI-PM, LOT NUMBER UNKNOWN