SERVO-I
Report
- Report Number
- 8010042-2011-00106
- Event Type
- Malfunction
- Date Received
- June 17, 2011
- Date of Event
- April 11, 2009
- Report Date
- May 20, 2011
- Manufacturer
- MAQUET CRITICAL CARE AB
- Product Code
- CBK
- PMA / PMN Number
- K062793
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE REPORTED AIR GAS MODULE WAS NOT RETURNED FOR INVESTIGATION. THE EVAL OF THE VENTILATOR LOGS SHOWED THAT THE PRESSURE TRANSDUCER TEST AND THE SAFETY VALVE TEST FAILED DURING PRE-USE CHECK. BASED ON THE EVAL OF THE LOGS, THE MOST PROBABLE CAUSE OF THESE FAILURES WAS A STICKY MEMBRANE IN THE NOZZLE UNIT MOUNTED IN THE AIR GAS MODULE. THE CAUSE OF THE STICKINESS IS WEAR OF THE MEMBRANE. THE REPORTED INTERNAL LEAKAGE TEST FAILURE COULD NOT BE CONFIRMED IN THE VENTILATOR LOGS. THE NOZZLE UNIT, WHICH IS PART OF THE GAS MODULE AND CONSISTS OF A MEMBRANE, A MOUTHPIECE AND A FEATHER SPRING, IS USED FOR REGULATION OF THE GAS FLOW THROUGH THE GAS MODULE. THE MEMBRANE IN THE NOZZLE UNIT IS PRESSED AGAINST A MOUTHPIECE WITH A FEATHER SPRING TO REGULATE THE GAS FLOW THROUGH THE GAS MODULE. THE MEMBRANE STICKINESS MAY CAUSE THE FEATHER SPRING TO GET STUCK TO THE MEMBRANE WHILE BEING PRESSED ONTO IT AND THUS CAUSING A DELAY ON RELEASE. THE DELAY AFFECTS THE REGULATION OF THE GAS FLOW THROUGH THE GAS MODULE AND CAUSES FAILURE DURING PRE-USE CHECK AND, DURING VENTILATION, CAUSES THE DELIVERED GAS FLOW AND PRESSURE TO BE HIGHER THAN SET. A FEATHER SPRING WITH A COARSE SURFACE WAS INTRODUCED IN (B)(6) 2010 TO PREVENT IT FROM GETTING STUCK TO THE MEMBRANE. (B)(4).
IT WAS REPORTED THAT THE VENTILATOR FAILED THE INTERNAL LEAKAGE TEST DURING PRE-USE CHECK. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SERVO-I | CBK | MAQUET CRITICAL CARE AB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |