MANUAL ORTHOPEDIC SURGICAL INSTRUMENT
Report
- Report Number
- 1030489-2011-01102
- Event Type
- Malfunction
- Date Received
- August 25, 2011
- Date of Event
- July 25, 2011
- Report Date
- September 14, 2011
- Manufacturer
- MEDTRONIC SOFAMOR DANEK INSTRUMENT MANUFACTURING
- Product Code
- LXH
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- OTHER
Narratives
VISUALLY CONFIRMED APPROXIMATELY ~3MM OF BOTH INSTRUMENT TIPS HAVE BEEN BROKEN OFF, CONSISTENT WITH INTERFACE DURING USAGE. DIMENSIONAL INSPECTION OF THE RELEVANT DIMENSION CONFIRMED CONFORMANCE TO PRINT SPECIFICATION. FRACTURE SURFACE ANALYSIS REVEALS A FAIRLY FLAT FRACTURE SURFACE AND CIRCULAR MATERIAL FLOW, CONSISTENT WITH TORSIONAL OVERLOAD, WITH THE TIP JUST BELOW THE FRACTURE SURFACES ARE PLASTICALLY DEFORMED.
(B)(4). NEITHER THE DEVICE NOR APPLICABLE IMAGING FILMS WERE RETURNED TO THE MANUFACTURER FOR EVALUATION, THEREFORE, THE CAUSE OF THE EVENT CANNOT BE DETERMINED. DEVICE HISTORY RECORDS FOR THESE LOTS WERE REVIEWED. NO DOCUMENTATION WAS FOUND THAT WOULD INDICATE A NON-CONFORMANCE TO SPECIFICATIONS.
IT WAS REPORTED THAT THE TIP BROKE DURING SURGERY. NO PATIENT COMPLICATIONS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MANUAL ORTHOPEDIC SURGICAL INSTRUMENT | ORTHOPEDIC MANUAL SURGICAL INSTRUMENT | LXH | MEDTRONIC SOFAMOR DANEK INSTRUMENT MANUFACTURING | NA | M06D0264 |