FDA Adverse Event Malfunction Summary report: N

EXTRA LONG STRAIGHT SABER ATTACHMENT

MDR report key: 2223268 · Received August 16, 2011

Report

Report Number
1811755-2011-03012
Event Type
Malfunction
Date Received
August 16, 2011
Date of Event
July 28, 2011
Report Date
July 28, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DRILL ATTACHMENT WAS RETURNED TO THE MANUFACTURER AND THE COMPLAINT WAS CONFIRMED. INTERNAL COMPONENTS WERE EVALUATED, WHICH REVEALED CORRODED BEARINGS AND FOREIGN DEBRIS CONTAMINATION. THE PRESENCE OF CORROSION AND DEBRIS CAN LEAD TO INCREASED FRICTION AMONG ROTATING COMPONENTS, RESULTING IN HEAT BEING GENERATED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TESTING CONDUCTED BY A MANUFACTURER FIELD SERVICE TECHNICIAN, THE DRILL ATTACHMENT WAS GETTING OVER-HEATED. THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE, SO THERE WAS NO PATIENT INVOLVEMENT. NO USER INJURY WAS REPORTED, AND NO OTHER ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EXTRA LONG STRAIGHT SABER ATTACHMENT ERL STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK