FDA Adverse Event Malfunction Summary report: N

STRAIGHT MEDIUM ELITE ATTACHMENT

MDR report key: 2223263 · Received August 16, 2011

Report

Report Number
1811755-2011-03011
Event Type
Malfunction
Date Received
August 16, 2011
Date of Event
August 5, 2011
Report Date
August 5, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
HBE
PMA / PMN Number
K032303
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DRILL ATTACHMENT WAS RETURNED TO THE MANUFACTURER AND THE COMPLAINT WAS CONFIRMED. INTERNAL COMPONENTS WERE EVALUATED, WHICH REVEALED GRITTY BEARINGS, CORROSION AND FOREIGN DEBRIS CONTAMINATION. THE PRESENCE OF CORROSION AND DEBRIS CAN LEAD TO INCREASED FRICTION AMONG ROTATING COMPONENTS, RESULTING IN HEAT BEING GENERATED.

Description of Event or Problem · 1

FIELD SERVICE TECHNICIAN STATED THAT THE PRODUCT OVER-HEATED DURING A ROUTINE PROCARE VISIT. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE CONSEQUENCES NOTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STRAIGHT MEDIUM ELITE ATTACHMENT HBE STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK