UNICEL DXC 600 SYNCHRON CLINICAL SYSTEM
Report
- Report Number
- 2050012-2011-04421
- Event Type
- Malfunction
- Date Received
- August 12, 2011
- Date of Event
- April 8, 2008
- Report Date
- April 9, 2008
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
Narratives
THIS IS REPORT 31 OF 35 MEDWATCH REPORTS FILED FOR THIS EVENT FOR PT 31 OF 35 PTS. SERVICE WAS NOT REQUESTED. THE CUSTOMER WAS PROVIDED WITH THE APPROVED ALTERNATE FLOW CELL MAINTENANCE CLEANING PROCEDURE. THE CUSTOMER STATED THAT THEY HAVE INCREASED THE QC RUN FREQUENCY TO EVERY FOUR HOURS. THE EVENT APPEARS TO BE RELATED TO THE WEEKLY FLOWCELL MAINTENANCE CLEANING, BUT A CLEAR ROOT CAUSE HAS NOT BEEN DETERMINED. THIS REPORTABLE EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW OF COMPLAINTS CONDUCTED BETWEEN (B)(4) 2008 AND (B)(4) 2010 FOR ADD'L REPORTABLE EVENTS.
THE CUSTOMER CONTACTED BECKMAN COULTER, INC. (BCI) TO REPORT THAT THEIR UNICEL DXC 600 SYNCHRON CLINICAL SYSTEM GAVE ERRONEOUSLY LOW SODIUM (NA) AND/OR CALCIUM (CALC) RESULTS FOR MULTIPLE PTS. THE ERRONEOUS RESULTS WERE REPORTED OUT OF THE LAB. IT IS UNK IF THERE WERE ANY CHANGES TO PT TREATMENT ASSOCIATED WITH THIS EVENT, BUT THERE HAVE BEEN NO REPORTS OF DEATH OR SERIOUS INJURY. THE CUSTOMER STATED THAT THE WEEKLY AUTOMATED FLOWCELL MAINTENANCE CLEANING PROCEDURE WAS COMPLETED AT APPROX 1:00PM. THE ION SELECTIVE ELECTRODES (ISES) WERE CALIBRATED AND A QUALITY CONTROL (QC) WAS RUN WITH GOOD RESULTS. AT APPROX 10:00PM, THE CUSTOMER WAS ALERTED TO TWO CRITICALLY LOW NA AND CALC RESULTS. ALL RESULTS RUN SINCE THE MAINTENANCE WERE REVIEWED AND IT WAS NOTED THAT NA AND CALC STARTED TO RECOVER LOW SHORTLY AFTER THE MAINTENANCE WAS COMPLETED. ALL SAMPLES RUN SINCE THE MAINTENANCE WERE REPEATED ON THE LAB'S SECOND DXC ANALYZER. THE NA AND CALC RESULTS WERE HIGHER. THE CUSTOMER STATED THAT APPROX 35 CORRECTED REPORTS WERE ISSUED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL DXC 600 SYNCHRON CLINICAL SYSTEM | JJE | BECKMAN COULTER, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |