FDA Adverse Event Malfunction Summary report: N

ADJ PIN COLLET 2.0-3.2MM

MDR report key: 2223213 · Received August 16, 2011

Report

Report Number
1811755-2011-03023
Event Type
Malfunction
Date Received
August 16, 2011
Date of Event
August 5, 2011
Report Date
August 5, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
KIJ
PMA / PMN Number
K972367
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DE, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR INVESTIGATION. BASED ON THE QUALITY INVESTIGATION, THE IMPREGLON COATING ON THE DEVICE WAS WORN OFF. THIS CONDITION COULD CAUSE THE DEVICE TO NOT RETAIN THE PIN.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WOULD NOT RETAIN A PIN DURING ROUTINE MAINTENANCE. THERE WAS NO PATIENT INVOLVEMENT AND NO ALLEGATION OF ADVERSE CONSEQUENCES ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADJ PIN COLLET 2.0-3.2MM KIJ STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK