CENTRA BED
MDR report key: 2223144
·
Received August 16, 2011
Report
- Report Number
- 1824206-2011-04339
- Event Type
- Malfunction
- Date Received
- August 16, 2011
- Date of Event
- August 5, 2011
- Report Date
- August 5, 2011
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN REPLACED THE POWER CORD PLUG TO REPAIR THE BED.
Description of Event or Problem · 1
INFORMATION RECEIVED INDICATES THE TECHNICIAN FOUND THE BED HAS A HIGH GROUND READING DUE TO A BROKEN GROUND PIN ON THE AC POWER CORD PLUG.