Additional Manufacturer Narrative · 1

THE TECHNICIAN REPLACED THE POWER CORD PLUG TO REPAIR THE BED.

Description of Event or Problem · 1

INFORMATION RECEIVED INDICATES THE TECHNICIAN FOUND THE BED HAS A HIGH GROUND READING DUE TO A BROKEN GROUND PIN ON THE AC POWER CORD PLUG.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CENTRA BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 850