FDA Adverse Event
Malfunction
Summary report: N
E SERIES DEFIBRILLATOR
MDR report key: 2223130
·
Received August 17, 2011
Report
- Report Number
- 1220908-2011-02242
- Event Type
- Malfunction
- Date Received
- August 17, 2011
- Report Date
- July 27, 2011
- Manufacturer
- ZOLL MEDICAL CORPORATION
- Product Code
- MKJ
- PMA / PMN Number
- K042007
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORPORATION; THE MALFUNCTION WAS DUPLICATED AND ATTRIBUTED TO AN UNSEATED FLEX CABLE. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.
Description of Event or Problem · 1
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE POWERED UP WITHOUT DISPLAY. COMPLAINANT INDICATED THAT THERE WAS NO PT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | E SERIES DEFIBRILLATOR | E SERIES | MKJ | ZOLL MEDICAL CORPORATION | E SERIES | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |