ION PACLITAXEL-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM
Report
- Report Number
- 2134265-2011-03551
- Event Type
- Injury
- Date Received
- August 25, 2011
- Date of Event
- July 30, 2011
- Report Date
- July 30, 2011
- Manufacturer
- BOSTON SCIENTIFIC - MAPLE GROVE
- Product Code
- NIQ
- PMA / PMN Number
- P100023
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE IS A COMBINATION PRODUCT. PATIENT AGE AT TIME OF EVENT: 50S. (B)(4) - THE COMPLAINT DEVICE WAS NOT RECEIVED AT THE COMPLAINT INVESTIGATION SITE (CIS) FOR ANALYSIS. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).
SAME CASE AS MFR ID # 2134265-2011-03555. IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY TWO STENT DISLODGEMENTS OCCURRED. THE TARGET LESION WAS LOCATED IN THE RIGHT CORONARY ARTERY (RCA). THE 2.5X28MM ION STENT WAS ADVANCED TO THE RCA BUT WAS UNABLE TO CROSS A PREVIOUSLY PLACED STENT. THE PHYSICIAN ATTEMPTED TO PULL BACK THE STENT DELIVERY SYSTEM (SDS) AND NOTICED THAT THE STENT HAD MOVED ON THE BALLOON. THE PHYSICIAN REMOVED THE SDS WITH THE GUIDEWIRE, AND THE STENT DISLODGED IN THE SHEATH. THE SHEATH WAS REMOVED FROM THE PATIENT AND THE STENT WAS RECOVERED FROM THE SHEATH. THE 4.0X8MM ION STENT WAS ADVANCED TO THE RCA AND THE STENT DISLODGED PRIOR TO DEPLOYMENT. THE PHYSICIAN LOCATED THE AND CRUSHED THE EMBOLIZED STENT WITH A PROMUS STENT IN THE POSTERIOR DESCENDING ARTERY (PDA). PROCEDURE WAS COMPLETED WITH A PROMUS STENT. NO ADDITIONAL PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ION PACLITAXEL-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM | CORONARY DRUG-ELUTING STENT | NIQ | BOSTON SCIENTIFIC - MAPLE GROVE | H7493902408400 | 13984494 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |