MDR report key: 2223085 · Received August 10, 2011

Report

Report Number
1720753-2011-20899
Event Type
Malfunction
Date Received
August 10, 2011
Date of Event
August 2, 2011
Report Date
August 10, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE GE SERVICE REPRESENTATIVE CONDUCTED AN ONSITE INVESTIGATION. THE REPORTED PROBLEM COULD NOT BE DUPLICATED. THE COIN BATTERY WAS REPLACED. THE CALIBRATION FILES WERE LOADED AND THE CINE DRIVE WAS FORMATTED. THE GROUNDING PIN IN THE PLUG WAS TIGHTENED. THE SYSTEM WAS TESTED AND OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM INTERMITTENTLY WOULD NOT DISPLAY AN IMAGE. NO PATIENT INJURY WAS REPORTED.