- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE GE SERVICE REPRESENTATIVE CONDUCTED AN ONSITE INVESTIGATION. THE POWER SUPPLY WAS REPLACED AND THE VOLTAGE WAS ADJUSTED. THE SYSTEM WAS TESTED AND OPERATES AS INTENDED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE FLUOROSCOPY BUTTON WAS STUCK. NO PATIENT INJURY WAS REPORTED.