FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2223076 · Received August 10, 2011

Report

Report Number
1720753-2011-20907
Event Type
Malfunction
Date Received
August 10, 2011
Date of Event
August 3, 2011
Report Date
August 10, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE GE SERVICE REPRESENTATIVE CONDUCTED AN ONSITE INVESTIGATION. THE X-RAY CONTROLLER BOARD WAS RESEATED. THE SYSTEM WAS TESTED AND OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WOULD SHUT DOWN WITHOUT USER INPUT. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1