FDA Adverse Event
Other
Summary report: N
MEDICHOICE WARM PACK
MDR report key: 2223072
·
Received August 21, 2011
Report
- Report Number
- MW5021847
- Event Type
- Other
- Date Received
- August 21, 2011
- Date of Event
- April 8, 2011
- Report Date
- August 21, 2011
- Manufacturer
- COLDSTAR INTL INC
- Product Code
- IMD
- Report Source
- Voluntary report
- Reporter Location
- FL, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
WHILE ATTEMPTING TO REPLACE IV ACCESS ON AN INFANT, A HEATING PACK WAS PLACED DIRECTLY ON THE INFANT'S LEFT ARM. WHEN THE PACK WAS REMOVED LESS THAN 3 MINUTES LATER, THE INFANT'S SKIN WAS BLISTERED IN MULTIPLE AREAS AND SUFFERED A SECOND-DEGREE BURN. THE PACK WARMS TO 120 DEGREES, RECOMMENDS WRAPPING IN A TOWEL OR PROVIDED CLOTH SLEEVE PRIOR TO APPLICATION, AND STATES "DO NOT USE ON INFANTS." NO CLOTH SLEEVE WAS PROVIDED, AND THE 120 DEGREE PACK WAS THE SAME COLOR (RED AND WHITE STRIPED) AS ANOTHER, SLIGHTLY SMALLER, PACK BY ANOTHER MFR WHICH HEATS TO 110 DEGREES, DOES NOT REQUIRE A SLEEVE OR TOWEL, AND HAS NO INFANT RESTRICTIONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MEDICHOICE WARM PACK | WARM PACK | IMD | COLDSTAR INTL INC | 69901 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 2 MO | Other | CATALOG #11443-512| CARDINAL HEALTH "INSTANT HOT PACK": |