FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2223040 · Received August 10, 2011

Report

Report Number
1720753-2011-20895
Event Type
Malfunction
Date Received
August 10, 2011
Date of Event
July 22, 2011
Report Date
August 10, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SVC REP PERFORMED AN ONSITE INVESTIGATION. THE SVC REP REPLACED THE CENTRAL PROCESSING UNIT AND THE SYSTEM HARD DISK DRIVE AND THE INTERCONNECT CABLE. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK IN SVC.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WOULD DISPLAY A HARD DISK FAILURE ERROR MESSAGE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1