MDR report key: 2223020 · Received August 10, 2011

Report

Report Number
1720753-2011-20925
Event Type
Malfunction
Date Received
August 10, 2011
Date of Event
August 5, 2011
Report Date
August 10, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE CONNECTION ON THE POWER SIGNAL INTERFACE BOARD WAS RESEATED WHICH INCREASED THE VOLTAGE ON THE HIGH VOLTAGE PRINTED CIRCUIT BOARD. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT DURING A PROCEDURE THE SYSTEM DISPLAYED A COMMUNICATION ERROR MESSAGE AND WAS IN CONTINUOUS FLUOROSCOPY MODE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1