FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 2222996
·
Received August 10, 2011
Report
- Report Number
- 1720753-2011-20931
- Event Type
- Malfunction
- Date Received
- August 10, 2011
- Date of Event
- July 21, 2011
- Report Date
- August 10, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE HARD DRIVE WAS REPLACED AND THE SOFTWARE WAS RELOADED. THE ROLLER ROCKER KIT WAS INSTALLED AND THE C-ARM WAS REMOVED AND RE-INSTALLED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT SAVE INFO AND DISPLAYED A CHECK FILE ERROR MESSAGE, AND WOULD NOT BOOT UP. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |