TLS3 23C
Report
- Report Number
- 2954339-2011-00018
- Event Type
- Malfunction
- Date Received
- August 10, 2011
- Date of Event
- July 19, 2011
- Report Date
- August 9, 2011
- Manufacturer
- STARION INSTRUMENTS
- Product Code
- GEI
- PMA / PMN Number
- K062257
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
ONE TLS3 23C WAS RETURNED, DECONTAMINATED AND INVESTIGATED. THE SERIAL NUMBER COULD NOT BE DETERMINED DUE TO THE CORD BEING CUT ON THE RETURNED TLS3 23C. A VISUAL INSPECTION OF THE RETURNED DEVICE WAS PERFORMED. THE RIGHT HAW BOOT WAS TORN ON THE OUTSIDE SURFACE, CONFIRMING THE COMPLAINT. THE TORN BOOT PIECE WAS NOT RETURNED. THERE ALSO WAS SUBSTANTIAL CHARRED TISSUE ATTACHED TO THE JAW BOOT. THE GAP BETWEEN THE HEAT SPREADER AND THE JAW WITH THE TORN BOOT WAS APPROPRIATE, NO DEFECTS WERE VISIBLE. THE BOOT TEAR IS CONSISTENT WITH JAWS THAT ARE NOT COMPLETELY CLOSED DURING INSERTION OR WITHDRAWAL OF THE DEVICE THROUGH A CANNULA. INCOMPLETE CLOSURE DURING CANNULA INSERTION AND/OR WITHDRAWAL CAN CAUSE THE JAWS TO BE SLIGHTLY APART FORCING THEM TO COME IN CONTACT WITH THE WALLS OF THE CANNULA AND THE BOOT AND OTHER COMPONENTS CAN GET DAMAGED. TEARS MAY ENLARGE WITH EXTENSIVE USE OR AGGRESSIVE CLEANING. NO ISSUES WERE FOUND WITH THE DEVICE HISTORY RECORD.
DURING A RADICAL PROSTATECTOMY PROCEDURE, THE SCRUB TECH WIPED THE JAWS TO CLEAN THEM AND A SMALL PIECE OF THE VERY END OF THE SILICONE TIP CAME OFF. THIS DID NOT HAPPEN INSIDE THE PT, HOWEVER, IT WAS DURING THE SURGERY. NO OTHER PT INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TLS3 23C | LIGATING SHEAR 35 C | GEI | STARION INSTRUMENTS | 132-135D | 902003 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |