FDA Adverse Event Malfunction Summary report: N

ADVANTA BED

MDR report key: 2222875 · Received July 15, 2011

Report

Report Number
1824206-2011-03747
Event Type
Malfunction
Date Received
July 15, 2011
Date of Event
June 29, 2011
Report Date
June 29, 2011
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECH REPLACED THE POWER BOARD TO REPAIR THE BED.

Description of Event or Problem · 1

INFO RECEIVED INDICATES THE POWER BOARD HAS SOME CORROSION ON IT DUE TO WHAT APPEARS TO BE LIQUID OF SOME KIND.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADVANTA BED AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1604

Patients

Seq Age Sex Outcome Treatment
1