HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1423500-2011-11240
- Event Type
- Malfunction
- Date Received
- August 25, 2011
- Date of Event
- August 1, 2011
- Report Date
- August 1, 2011
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THIS COMPLAINT FOR A SYSTEM ERROR 2240 (AIR IN SET) OCCURRED WAS NOT CONFIRMED; PER THE COMPLAINT INFORMATION THE CAUSE OF THE SYSTEM ERROR 2240 WAS DUE TO AIR BEING SUCKED INTO THE DISPOSABLE BECAUSE THE PATIENT LINE BECAME INADVERTENTLY DISCONNECTED FROM THE TRANSFER SET. THE ASSIGNABLE CAUSE WAS DUE TO USE ERROR - PATIENT DISCONNECTED FROM SET. A LABEL REVIEW WAS PERFORMED AND THE REVIEW FOUND THE LABELING ADEQUATE FOR THE USER ERROR IN THE COMPLAINT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS.
(B)(4). THE SAMPLE WAS NOT RETURNED TO BAXTER, THEREFORE NO EVALUATION OR BATCH REVIEW COULD BE PERFORMED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF BAXTER'S INVESTIGATION, OR IF ADDITIONAL INFORMATION IS RECEIVED.
A CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER REGARDING A SYSTEM ERROR 2240 (AIR IN LINE) ALARM THAT APPEARED ON THE HOMECHOICE (HC) DISPLAY DURING DRAIN 5 OF 6. THE HOME PATIENT (HP) REPORTEDLY FELL OUT OF THE BED AND THE LINE DISCONNECTED. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED THE ALARM TO THE HP'S SON AND ASSISTED TO CLEAR THE ALARM. THE HP WOULD STOP THERAPY UNTIL CONTACT WAS MADE WITH NURSE. THERE WAS PATIENT INVOLVEMENT. THERE WAS NO INJURY OR MEDICAL INTERVENTION REPORTED AS A RESULT OF THIS EVENT. DURING A FOLLOW UP PHONE CALL BY PRODUCT SURVEILLANCE TO THE CAREGIVER (CG) ON (B)(6) 2011 REGARDING THE ALARM, IT WAS REVEALED THAT THE ALARM OCCURRED DUE TO A DISCONNECTION. THE CG STATED THERE WERE NO DEFECTS ON THE SUPPLIES. THE CG STATED THE ALARM OCCURRED BECAUSE OF THE WAY THEY DISCONNECTED THE HP. THE CG STATED THE HP HAD NO INJURY OR MEDICAL INTERVENTION FROM THIS INCIDENT. THE CG COULD NOT PROVIDE FURTHER DETAILS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR | HOMECHOICE |