TALENT THORACIC STENT GRAFT SYSTEM
Report
- Report Number
- 2953200-2011-01435
- Event Type
- Death
- Date Received
- August 12, 2011
- Date of Event
- September 13, 2012
- Report Date
- September 11, 2013
- Manufacturer
- MEDTRONIC CARDIOVASCULAR GALWAY
- Product Code
- MIH
- PMA / PMN Number
- P070007
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). EVALUATION, RESULTS: INHERENT RISK OF PROCEDURE (RUPTURE); PATIENT¿S CONDITION AFFECTED EFFECTIVENESS OF DEVICE (ANATOMY RELATED; CONTINUE DISEASE PROGRESSION). CONCLUSION: DEVICE FAILURE/LACK OF EFFECTIVENESS RELATED TO PATIENT CONDITION (ANATOMY RELATED; CONTINUE DISEASE PROGRESSION).
(B)(4). EVAL, RESULTS: (ENDOLEAK), (IMPLANTING THE STENT GRAFT IN THE PROXIMAL THORACIC AORTA). CONCLUSION: (IMPLANTING THE STENT GRAFT IN THE PROXIMAL THORACIC AORTA).
ADDITIONAL INFORMATION WAS RECEIVED FOR THIS CASE. IT WAS REPORTED FROM THE PHYSICIAN THAT POST-OPERATIVE FOLLOW-UP CT DATA SHOWED AN ENDOLEAK. IT WAS ORIGINALLY DEEMED TO BE TYPE III EL, BUT IT WAS DETERMINED TO BE TYPE IA EL THROUGH SEVERAL TIMES CT CHECK TWO YEARS AGO. ONE YEAR AGO, THE PATIENT WAS TAKEN TO HOSPITAL DUE TO ANEURYSM RUPTURE BECAUSE OF EXPANSION OF ANEURYSM DIAMETER WHOSE SIZE WAS 7. 3 CM IN DIAMETER, DUE TO THE UNTREATED TYPE IA ENDOLEAK. THE RUPTURED ANEURYSM BLOOD FLOWED INTO BRONCHUS AND PATIENT HEMORRHAGED. THE PATIENT EXPIRED THE FOLLOWING DAY DUE TO ANEURYSM RUPTURE. AT THE TIME OF DEVICE IMPLANT, TOUCH-UP WAS GIVEN WITHOUT LEAK AND PROCEDURE WAS COMPLETED. THE ENDOLEAK WAS APPEARED AT 1 MONTH LATER FROM TEVAR, BUT THE CAUSE WAS UNKNOWN. THIS WAS THE CASE THAT DOCTOR HAD A TALK WITH THE PATIENT AND IT WAS CONSIDERED RISKS IN ADDITIONAL TREATMENT BECAUSE OF HIS OLD AGE AND HAVING OTHER DISEASES, THEN IT WAS DECIDED NOT TO HAVE ADDITIONAL TREATMENT, BUT TO HAVE CONSERVATIVE TREATMENT FOLLOWING HIS STRONG REQUEST. THE CAUSE OF DEATH WAS ANEURYSM RUPTURE. PATIENT CONDITION WAS NOT GOOD HEALTH AND PREFERRED TO HAVE CONSERVATIVE TREATMENT, SO THAT T HERE WAS NO OTHER EFFECTIVE MEANS.
A TALENT THORACIC STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF A FUSIFORM 6.2 CM IN DIAMETER X 10 CM LONG THORACIC AORTIC ANEURYSM IN ZONE 1 APPROX 9 MONTHS AGO. THE PROXIMAL AORTIC NECK DIAMETER WAS 42 MM, AND THE DISTAL AORTIC NECK WAS 34 MM IN DIAMETER. THERE WAS NO CALCIFICATION, BUT THERE WAS MILD THROMBUS AT THE DISTAL AORTIC NECK. IT WAS REPORTED THAT APPROX 3 WEEKS POST-IMPLANTATION, AN UNK ENDOLEAK, POSSIBLY A DISTAL TYPE I ENDOLEAK, A TYPE II ENDOLEAK, OR A JUNCTIONAL TYPE III ENDOLEAK BETWEEN THE DISTAL (B)(4) STENT GRAFT (MFR REPORT # 2953200-2011-01432) AND THE PROXIMAL (B)(4) STENT GRAFT (MFR REPORT # 2953200-2011-01433) WAS SEEN VIA CT. THE DIAMETER OF THE ANEURYSM HAS NOT BEEN INCREASING; THEREFORE, THE PHYSICIAN ELECTED TO MONITOR THE PT WITHOUT ANY INTERVENTION. ON AN UNK DATE, A PROXIMAL TYPE I ENDOLEAK WAS CONFIRMED, AND APPROX 1 MONTH AGO, THE PT RETURNED FOR AN INTERVENTION, AS THE PHYSICIAN PERFORMED COIL EMBOLIZATION OF THE LEFT SUBCLAVIAN ARTERY FOR A CONFIRMED TYPE II ENDOLEAK. AFTER THE EMBOLIZATION, THE PROXIMAL TYPE I ENDOLEAK WORSENED. THE PHYSICIAN ELECTED TO INTERVENE BY IMPLANTING A (B)(4) STENT GRAFT (MFR REPORT # 2953200-2011-01434) PROXIMALLY; HOWEVER, THE BARE SPRING INVERTED, AND THEN THE PHYSICIAN PULLED BACK ON THE DELIVERY SYSTEM TO CORRECT THE INVERTED BARE SPRING. ALTHOUGH THE BARE SPRING WAS CORRECTED, THE STENT GRAFT MOVED FROM ITS INTENDED PLACEMENT. THE PHYSICIAN ELECTED TO IMPLANT ANOTHER (B)(4) STENT GRAFT (MFR REPORT # 2953200-2011-01435); HOWEVER, THE BARE SPRING INVERTED, AND THE PHYSICIAN PUSHED AND PULLED ON THE DELIVERY SYSTEM TO TRY TO CONNECT THE BARE SPRING'S POSITION. THE PHYSICIAN WAS UNABLE TO CORRECT THE BARE SPRING, WHICH REMAINED MISALIGNED, AND THE PROXIMAL TYPE I ENDOLEAK REMAINED. NO FURTHER INTERVENTION WAS PERFORMED FOR THE PROXIMAL TYPE I ENDOLEAK. THERE WAS ALSO A DISTAL TYPE I ENDOLEAK; THE PHYSICIAN ELECTED TO INTERVENE BY IMPLANTING A (B)(4) STENT GRAFT DISTALLY, WHICH SUCCESSFULLY RESOLVED THE DISTAL TYPE I ENDOLEAK. NO ADD'L CLINICAL SEQUELAE WERE REPORTED, AND THE PT IS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TALENT THORACIC STENT GRAFT SYSTEM | MIH | MEDTRONIC CARDIOVASCULAR GALWAY | NA | V00541197 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR | Death |