FDA Adverse Event Summary report: N

12X150MM KII SHIELDED BLADED ACCESS SYS THREADED

MDR report key: 2222728 · Received July 8, 2011

Report

Report Number
2027111-2011-00064
Date Received
July 8, 2011
Date of Event
January 12, 2011
Report Date
July 8, 2011
Manufacturer
APPLIED MEDICAL
Product Code
GCJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION SUMMARY: THE DEVICE INVOLVED IN THE INCIDENT WAS NOT RETURNED FOR EVALUATION; HOWEVER, THE REPORT STATED THE DEVICE WAS USED WITH A DA VINCI SURGICAL ROBOT SYSTEM. THE C0R69 THREADED CANNULA MODEL IS NOT QUALIFIED FOR USE WITH THE DA VINCI ROBOT SYSTEM. APPLIED RECOMMENDS REVIEW OF THE INTUITIVE SURGICAL, INC.'S (DA VINCI ROBOT MANUFACTURER) INSTRUMENTS AND ACCESSORIES TO UNDERSTAND WHICH APPLIED MEDICAL TROCARS HAVE BEEN VALIDATED FOR USE WITH THE DA VINCI ROBOT AND THE CORRESPONDING ADAPTERS. INTUITIVE SURGICAL, INC. SUPPLIES A VALIDATED MOUNT (MODEL# 371528) FOR APPLIED MEDICAL NON-THREADED TROCAR MODELS. THIS DOCUMENT REPRESENTS OUR INITIAL AND FINAL REPORT. ADDITIONAL COMMENTS: THE CUSTOMER EXPERIENCE REPORT SUBMITTED TO APPLIED MEDICAL MADE NO REFERENCE TO PT INJURY OR REQUIRED INTERVENTION. FOLLOW UP WITH THE HOSPITAL CONFIRMED THERE WAS INTERVENTION TO ENSURE RECOVERY OF ANY PIECES OF THE DEVICE; HOWEVER, THERE WAS NO PT INJURY.

Description of Event or Problem · 1

ROBOTIC HYSTERECTOMY- "TROCAR CRACKED AND BROKE ON THE SIDE UNABLE TO KEEP PNEUMO PERITONEUM, 2 PIECES WERE BROKEN OFF AND FOUND ON THE OUTSIDE OF THE PTS ABDOMEN. TWO SEPARATE CRACKS STARTED NEAR THE SEAL HOUSING AND CONTINUED DOWN THE TROCAR. ONE OF THE CRACKS HAD A PIECE (AND COVERED 8 THREADS) BREAK OFF THAT WAS "CRESCENT SHAPED."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 12X150MM KII SHIELDED BLADED ACCESS SYS THREADED NONE GCJ APPLIED MEDICAL C0R69 1118989

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention DA VINCI ROBOT CAMERA