FDA Adverse Event Malfunction Summary report: N

STERLING¿

MDR report key: 2222726 · Received August 25, 2011

Report

Report Number
2134265-2011-03585
Event Type
Malfunction
Date Received
August 25, 2011
Date of Event
August 4, 2011
Report Date
August 5, 2011
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
DQY
PMA / PMN Number
K053116
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR: THE F/G STERLING OTW BALLOON CATHETER WAS RECEIVED FOR ANALYSIS. WHEN POSITIVE PRESSURE WAS APPLIED WITH AN INFLATION DEVICE FILLED WITH WATER, A STREAM OF WATER EMITTED FROM A PINHOLE IN THE MIDSECTION OF THE BALLOON. VISUAL INSPECTION OF THE BALLOON PRESENTED NO IRREGULARITIES IN THE BALLOON MATERIAL OR THE RO MARKERS THAT COULD HAVE CONTRIBUTED TO THE DAMAGE. THERE WAS NO EVIDENCE OF ANY PRODUCT QUALITY DEFICIENCIES CONTRIBUTING TO THE DAMAGE. THE MANUFACTURING RECORD CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT 18 YEARS OR OLDER. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY TREATMENT PROCEDURE A BALLOON RUPTURE OCCURRED. THE 90% STENOSED LESION WAS LOCATED IN THE MODERATELY TORTUOUS LEFT CUBITAL REGION. THE 6.0 X 40/80 F/G STERLING OTW BALLOON CATHETER RUPTURED AT 14ATMS ON THE INITIAL INFLATION. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS GOOD.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY TREATMENT PROCEDURE A BALLOON RUPTURE OCCURRED. THE 90% STENOSED LESION WAS LOCATED IN THE MODERATELY TORTUOUS LEFT CUBITAL REGION. THE 6.0 X 40/80 F/G STERLING OTW BALLOON CATHETER RUPTURED AT 14ATMS ON THE INITIAL INFLATION. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STERLING¿ CATHETER, PERCUTANEOUS DQY BOSTON SCIENTIFIC - MAPLE GROVE H74939032604080 13821773

Patients

Seq Age Sex Outcome Treatment
1