FDA Adverse Event
Other
Summary report: N
GENESIS IPG
MDR report key: 2222537
·
Received August 11, 2011
Report
- Report Number
- 1627487-2011-02913
- Event Type
- Other
- Date Received
- August 11, 2011
- Date of Event
- May 23, 2011
- Report Date
- May 23, 2011
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
EVALUATION CONCLUSION: THE LOW BATTERY WARNING WAS DUE TO PASSIVATION. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
Description of Event or Problem · 1
THE PATIENT RECEIVED HIS SCS SYSTEM ON (B)(6) 2010. IT WAS REPORTED THAT THE IPG WAS EXPLANTED AND REPLACED ON (B)(6) 2011 DUE TO BATTERY DEPLETION. THE EXPLANTED IPG WAS RETURNED TO THE MANUFACTURER FOR ANALYSIS. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GENESIS IPG | TOTALLY IMPLANTABLE PULSE GENERATOR | LGW | ST. JUDE MEDICAL - NEUROMODULATION | 3608 | 3165993 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | Required Intervention | SCS LEAD: MODEL 3186| IMPLANT: |