FDA Adverse Event Other Summary report: N

SMARTPLUG

MDR report key: 2222483 · Received April 26, 2011

Report

Report Number
2031959-2011-00001
Event Type
Other
Date Received
April 26, 2011
Date of Event
January 1, 2002
Report Date
April 25, 2011
Manufacturer
MEDENNIUM, INC.
Product Code
LZU
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

FROM THE DATES IDENTIFIED ON THE ARTICLE WE ONLY SHOW SHIPPING RECORDS FROM 2002 TO 2005 OF A TOTAL OF (B)(4). IMPORTANT TO NOTE THAT NO MEDICAL TREATMENT WAS ATTEMPTED ON 6 OF THE 17 PATIENTS SO THE STATEMENT SHOULD READ ALL AFFECTED PATIENTS WERE TREATED BY CANALICULOTOMY AS THE TREATMENT OF CHOICE. THE LONG TERM STUDIES SUGGEST 3.9% FOR SILICONE HEADED PLUGS. CHEN SX, LEE GA 2007 REFERENCE NO. 10 IN THE HILL ARTICLE. SO IS THERE A SUBSTANTIAL DIFFERENCE IN THE RISKS OF CANALICULITIS WITH EITHER HEADED SILICONE PLUGS OR INTRACANALICULAR SMARTPLUGS? 3.9% SILICONE VS. 1.86% SMARTPLUG? REFERENCE NUMBER 7. THE SMARTPLUG STUDY GROUP HEADED BY JOSEPH MAURIELLO M.D. IN THIS ARTICLE AMONG 28 PATIENTS WITH DIAGNOSED COMPLICATIONS 9 OF THEM WERE SUCCESSFULLY TREATED WITHOUT THE NEED FOR SURGERY. THIS IS ABOUT 33%. IS THE BERSANI/HILL APPROACH TO RESOLVING COMPLICATIONS THE RIGHT APPROACH? IS IT TO PERFORM SURGERY WITHOUT ATTEMPTING TO RESOLVE THE SITUATION BY THE USE OF ANTIBIOTICS ORAL/TOPICAL WAITING FOR RESOLUTION OF INFECTION AND THEN ATTEMPTING IRRIGATION? ALSO IMPORTANT TO NOTE THAT BEFORE GIVING UP ON ANTIBIOTIC TREATMENT IT MAY BE HELPFUL TO STUDY IF THE ANTIBIOTIC BEING USED IS THE CORRECT ONE FOR THE TYPE OF BACTERIA BEING TREATED. THE PATIENTS HERE PRESENTED HAVE RESOLVED THEIR CANALICULITIS AND NO FURTHER REPORTS CAN BE EXPECTED. THIS REPORT IS CLOSED.

Description of Event or Problem · 1

(B)(4), MD ET ALL., "PREVALENCE OF CANALICULITIS REQUIRING SURGICAL REMOVAL OF SMART PLUGS", OPHTHALMIC PLASTIC RECONSTRUCTIVE SURGERY, VOL 25, NO6, 2009 WAS WRITTEN REGARDING SMART PLUG AND THE PREVALENCE OF CANALICULITIS IN ONE PRACTICE AS (B)(4). THE AVERAGE ONSET OF SYMPTOMS WAS AT 3 YEARS AND IN 3 CASES IN AS LITTLE AS 3, 4 AND 5 MONTHS, RESPECTIVELY. THE PLUGS WERE INSERTED FROM 2002 TO 2007. SEVENTEEN PATIENTS HAD SMART PLUG INSERTED AGES (B)(6), 15 FEMALE AND 2 MALE. TWO WERE IRRIGATED AND 7 RECEIVED TOPICAL ANTIBIOTICS. ONE RECEIVED TOPICAL AND ORAL ANTIBIOTICS. ONE RECEIVED ORAL ANTIBIOTICS ONLY. NO MEDICAL TREATMENT WAS ATTEMPTED IN 6 OF THE PATIENTS. CANALICULOTOMY WAS PERFORMED ON ALL PATIENTS. ONE PATIENT REQUIRED BILATERAL JONES TUBES DUE TO CHRONIC EPIPHORA. ONE PATIENT REQUIRED SILICONE INTUBATION FOR EPIPHORA. THE REPORTED OUTCOMES WERE THAT 9 RESOLVED, 4 HAD DRY EYE, 2 HAVING EPIPHORA ONE REQUIRING SILICONE INTUBATION AND ONE REQUIRING JONES TUBES. TWO HAD NO OUTCOMES LISTED ONE WITH NO MEDICAL TREATMENT ATTEMPTED AND ONE WITH TOPICAL ANTIBIOTICS ATTEMPTED. IT APPEARS THAT ALL PATIENTS WITH LISTED OUTCOMES HAD RESOLUTION OF THE CANALICULITIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SMARTPLUG PUNCTUM PLUG LZU MEDENNIUM, INC. 500 SEVERAL

Patients

Seq Age Sex Outcome Treatment
1 64 YR Required Intervention