FDA Adverse Event Malfunction Summary report: N

AFFINITY 4

MDR report key: 2222442 · Received July 14, 2011

Report

Report Number
1824206-2011-03741
Event Type
Malfunction
Date Received
July 14, 2011
Date of Event
June 30, 2011
Report Date
June 30, 2011
Manufacturer
HILL-ROM, INC.
Product Code
HDD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECH FOUND THE BRAKE WAS STICKING AND HE HAD TO MOVE PEDAL AROUND TO GET THE BRAKE TO DISENGAGE. HE FOUND THE BRAKE DETENT WAS CRACKED. HE REPLACED THE BRAKE DETENT TO REPAIR THE BED.

Description of Event or Problem · 1

INFO RECEIVED INDICATES THE BRAKE WILL NOT DISENGAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AFFINITY 4 BIRTHING BED HDD HILL-ROM, INC. 3700B

Patients

Seq Age Sex Outcome Treatment
1