FDA Adverse Event Malfunction Summary report: N

HEARTSTART MRX - EMS DEFIBRILLATOR

MDR report key: 2222408 · Received July 15, 2011

Report

Report Number
1218950-2011-02037
Event Type
Malfunction
Date Received
July 15, 2011
Report Date
June 20, 2011
Manufacturer
PHILIPS HEALTHCARE - ANDOVER
Product Code
MKJ
PMA / PMN Number
K031187
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). UNIT DOES NOT POWER ON WITH BATTERY IN EITHER BAY. NO REPORTED PATIENT INVOLVEMENT. THIS LOANER UNIT WAS EVALUATED AT THE PHILIPS REPAIR BENCH AND IT WAS FOUND THAT THE UNIT FAILED TO POWER UP ON BATTERY POWER DUE TO BROKEN BATTERY PCA CONNECTION PINS. THERE WAS NO PATIENT INVOLVEMENT. REPLACING THE BATTERY PCA RESOLVED THE FAILURE.

Description of Event or Problem · 1

UNIT DOES NOT POWER ON WITH BATTERY IN EITHER BAY. NO REPORTED PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HEARTSTART MRX - EMS DEFIBRILLATOR MKJ PHILIPS HEALTHCARE - ANDOVER M3536A

Patients

Seq Age Sex Outcome Treatment
1