FDA Adverse Event
Malfunction
Summary report: N
HEARTSTART MRX - EMS DEFIBRILLATOR
MDR report key: 2222408
·
Received July 15, 2011
Report
- Report Number
- 1218950-2011-02037
- Event Type
- Malfunction
- Date Received
- July 15, 2011
- Report Date
- June 20, 2011
- Manufacturer
- PHILIPS HEALTHCARE - ANDOVER
- Product Code
- MKJ
- PMA / PMN Number
- K031187
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
(B)(4). UNIT DOES NOT POWER ON WITH BATTERY IN EITHER BAY. NO REPORTED PATIENT INVOLVEMENT. THIS LOANER UNIT WAS EVALUATED AT THE PHILIPS REPAIR BENCH AND IT WAS FOUND THAT THE UNIT FAILED TO POWER UP ON BATTERY POWER DUE TO BROKEN BATTERY PCA CONNECTION PINS. THERE WAS NO PATIENT INVOLVEMENT. REPLACING THE BATTERY PCA RESOLVED THE FAILURE.
Description of Event or Problem · 1
UNIT DOES NOT POWER ON WITH BATTERY IN EITHER BAY. NO REPORTED PATIENT INVOLVEMENT.