FDA Adverse Event
Malfunction
Summary report: N
CONTOUR TEST STRIPS
MDR report key: 2222378
·
Received July 15, 2011
Report
- Report Number
- 1826988-2011-00425
- Event Type
- Malfunction
- Date Received
- July 15, 2011
- Date of Event
- June 1, 2011
- Report Date
- June 30, 2011
- Manufacturer
- BAYER HEALTHCARE, LLC
- Product Code
- NBW
- PMA / PMN Number
- K091820
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THE CUSTOMER TESTED HIS BLOOD GLUCOSE USING HIS CONTOUR USB METER AND RECEIVED A READING OF 122 MG/DL. HE RETESTED USING ANOTHER METER AND RECEIVED A READING OF 55 MG/DL. THE DIFFERENCE BETWEEN THE READINGS FALLS IN THE "C" ZONE OF THE CONSENSUS ERROR GRID, MAKING THE DIFFERENCE CLINICALLY SIGNIFICANT. NO ADVERSE EVENTS WERE ALLEGED. THE CUSTOMER WAS ADVISED TO RETURN HIS TEST STRIPS FOR EVAL. REPLACEMENT TEST STRIPS WERE SENT TO THE CUSTOMER. THE METER WAS ALSO REPLACED AT THE CUSTOMER'S INSISTENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONTOUR TEST STRIPS | NBW | BAYER HEALTHCARE, LLC | 7080G | 0JC3A09 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |