FDA Adverse Event Malfunction Summary report: N

CONTOUR TEST STRIPS

MDR report key: 2222367 · Received July 15, 2011

Report

Report Number
1826988-2011-00405
Event Type
Malfunction
Date Received
July 15, 2011
Date of Event
June 1, 2011
Report Date
June 21, 2011
Manufacturer
BAYER HEALTHCARE LLC
Product Code
NBW
PMA / PMN Number
K062058
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE CUSTOMER REC'D A BLOOD GLUCOSE READING OF 359 MG/DL FROM HER CONTOUR METER AND A READING OF 160 MG/DL FROM ANOTHER METER. THE DIFFERENCE BETWEEN THE READINGS FALLS IN THE "C" ZONE OF THE CONSENSUS ERROR GRID, MAKING THE DIFFERENCE CLINICALLY SIGNIFICANT. NO ADVERSE EVENT WAS ALLEGED. THE CUSTOMER WAS ADVISED TO RETURN THE TEST STRIPS FOR EVAL. REPLACEMENT STRIPS AND A NEW METER WERE SENT TO THE CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTOUR TEST STRIPS NBW BAYER HEALTHCARE LLC 7098B 9LC3D10

Patients

Seq Age Sex Outcome Treatment
1 UNK