SIEMENS BRAND NAME: ARTISTE MV
Report
- Report Number
- 2910081-2011-00037
- Event Type
- Other
- Date Received
- August 15, 2011
- Date of Event
- July 20, 2011
- Report Date
- July 20, 2011
- Manufacturer
- SIEMENS MEDICAL SOLUTIONS USA, INC.
- Product Code
- IYE
- PMA / PMN Number
- K072485
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
PRELIMINARY RISK ASSESSMENT INDICATES: SEVERITY: 3 (CRITICAL). THERE HAS BEEN NO MISTREATMENT OR INJURY REPORTED. THE SYSTEM WORKS AS SPECIFIED. THE PHYSICIST HAS TO ADAPT THE PLAN AND PERFORM THE MENTIONED CORRECTIONS OF THE DOSE BEFORE THE FIRST IRRADIATION IS POSSIBLE. WORST CASE: IF THE PHYSICIST MAKES A SEVERE MISTAKE WHEN MANUALLY ADAPTING THE MONITOR UNITS, IT IS POSSIBLE THAT AN INCORRECT DOSE WOULD BE DELIVERED FOR SEVERAL FRACTIONS. THIS COULD POTENTIALLY CAUSE A SERIOUS INJURY. PROBABILITY: B (IMPROBABLE) AFTER CHANGING THE PLAN, A NEW APPROVAL OF THE PLAN IS REQUIRED. IT IS VERY LIKELY, THAT A CONSIDERABLE DEVIATION FROM THE ORIGINALLY PLANNED DOSE WOULD BE RECOGNIZED AND CORRECTED BEFORE APPROVAL OF THE PLAN. NO OTHER PRODUCTS ARE CONCERNED AND FURTHER INVESTIGATION IS UNDERWAY FOR CAUSE DETERMINATION AND FINAL CORRECTIVE ACTION.
A POTENTIAL PRODUCT ISSUE HAS BEEN IDENTIFIED WITH OUR ARTISTE LINEAR ACCELERATOR. IN THE ABUNDANCE OF CAUTION, THIS PRODUCT PROBLEM IS BEING REPORTED. CUSTOMER USES A TREATMENT PLANNING SYSTEM, WHICH GIVES FRACTIONAL MONITOR UNITS FOR VIRTUAL WEDGE TREATMENTS, ELECTRON TREATMENTS AND ARC TREATMENTS. HOWEVER, RTT 4.2 DOES NOT SUPPORT FRACTIONAL MONITOR UNITS FOR THOSE ABOVE MENTIONED TREATMENTS. CUSTOMER IS UNABLE TO USE INTEGER MONITOR UNITS BECAUSE THEY REQUIRE FRACTIONAL MONITOR UNITS FOR STEP-AND-SHOOT INTENSITY MODULATED RADIATION THERAPY PLANS (IMRT). ACCORDING TO THE CUSTOMER, THE TREATMENT PLANNING SYSTEM DOES NOT ALLOW FRACTIONAL MONITOR UNITS FOR IMRT PLANS AND INTEGER MONITOR UNITS FOR OTHER PLANS. AS A RESULT, THE PHYSICIST HAS TO MANUALLY ADAPT THE PLAN AND PERFORM CORRECTIONS OF THE DOSE BEFORE THE FIRST IRRADIATION OF A PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SIEMENS BRAND NAME: ARTISTE MV | ACCELERATOR, LINEAR, MEDICAL | IYE | SIEMENS MEDICAL SOLUTIONS USA, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |