FDA Adverse Event Other Summary report: N

SIEMENS BRAND NAME: ARTISTE MV

MDR report key: 2222354 · Received August 15, 2011

Report

Report Number
2910081-2011-00037
Event Type
Other
Date Received
August 15, 2011
Date of Event
July 20, 2011
Report Date
July 20, 2011
Manufacturer
SIEMENS MEDICAL SOLUTIONS USA, INC.
Product Code
IYE
PMA / PMN Number
K072485
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

PRELIMINARY RISK ASSESSMENT INDICATES: SEVERITY: 3 (CRITICAL). THERE HAS BEEN NO MISTREATMENT OR INJURY REPORTED. THE SYSTEM WORKS AS SPECIFIED. THE PHYSICIST HAS TO ADAPT THE PLAN AND PERFORM THE MENTIONED CORRECTIONS OF THE DOSE BEFORE THE FIRST IRRADIATION IS POSSIBLE. WORST CASE: IF THE PHYSICIST MAKES A SEVERE MISTAKE WHEN MANUALLY ADAPTING THE MONITOR UNITS, IT IS POSSIBLE THAT AN INCORRECT DOSE WOULD BE DELIVERED FOR SEVERAL FRACTIONS. THIS COULD POTENTIALLY CAUSE A SERIOUS INJURY. PROBABILITY: B (IMPROBABLE) AFTER CHANGING THE PLAN, A NEW APPROVAL OF THE PLAN IS REQUIRED. IT IS VERY LIKELY, THAT A CONSIDERABLE DEVIATION FROM THE ORIGINALLY PLANNED DOSE WOULD BE RECOGNIZED AND CORRECTED BEFORE APPROVAL OF THE PLAN. NO OTHER PRODUCTS ARE CONCERNED AND FURTHER INVESTIGATION IS UNDERWAY FOR CAUSE DETERMINATION AND FINAL CORRECTIVE ACTION.

Description of Event or Problem · 1

A POTENTIAL PRODUCT ISSUE HAS BEEN IDENTIFIED WITH OUR ARTISTE LINEAR ACCELERATOR. IN THE ABUNDANCE OF CAUTION, THIS PRODUCT PROBLEM IS BEING REPORTED. CUSTOMER USES A TREATMENT PLANNING SYSTEM, WHICH GIVES FRACTIONAL MONITOR UNITS FOR VIRTUAL WEDGE TREATMENTS, ELECTRON TREATMENTS AND ARC TREATMENTS. HOWEVER, RTT 4.2 DOES NOT SUPPORT FRACTIONAL MONITOR UNITS FOR THOSE ABOVE MENTIONED TREATMENTS. CUSTOMER IS UNABLE TO USE INTEGER MONITOR UNITS BECAUSE THEY REQUIRE FRACTIONAL MONITOR UNITS FOR STEP-AND-SHOOT INTENSITY MODULATED RADIATION THERAPY PLANS (IMRT). ACCORDING TO THE CUSTOMER, THE TREATMENT PLANNING SYSTEM DOES NOT ALLOW FRACTIONAL MONITOR UNITS FOR IMRT PLANS AND INTEGER MONITOR UNITS FOR OTHER PLANS. AS A RESULT, THE PHYSICIST HAS TO MANUALLY ADAPT THE PLAN AND PERFORM CORRECTIONS OF THE DOSE BEFORE THE FIRST IRRADIATION OF A PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SIEMENS BRAND NAME: ARTISTE MV ACCELERATOR, LINEAR, MEDICAL IYE SIEMENS MEDICAL SOLUTIONS USA, INC.

Patients

Seq Age Sex Outcome Treatment
1