FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC SYNCHRON ® CHEMISTRY ANALYZER

MDR report key: 2222349 · Received August 25, 2011

Report

Report Number
2050012-2011-04702
Event Type
Malfunction
Date Received
August 25, 2011
Date of Event
July 25, 2011
Report Date
July 25, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SAMPLE INFORMATION AS NOT PROVIDED. THE CUSTOMER REPORTED TO CUSTOMER TECHNICAL SUPPORT (CTS) THAT THEY OBSERVED THE BUNM ELECTRODE HAD THE WRONG QUAD RING ATTACHED (DESCRIBED AS THINNER AND SMALLER BY CUSTOMER) AND IT WAS LEAKING. THE CUSTOMER REPLACED THE ELECTRODE WITH ONE THAT HAD THE CORRECT QUAD RING. CTS SENT THE CUSTOMER A REPLACEMENT ELECTRODE. THE CUSTOMER DID NOT GENERATED SERVICE REQUEST.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) TO REPORT ERRONEOUS BUNM RESULTS GENERATED BY THE UNICEL DXC 800 PRO SYNCHRON CHEMISTRY ANALYZER. NO ERRONEOUS RESULTS WERE REPORTED OUT OF THE LABORATORY. THE ERRONEOUS SAMPLES WERE REPEATED ON ANOTHER INSTRUMENT; HOWEVER, THE CUSTOMER WAS UNABLE TO PROVIDE DATA. PATIENT TREATMENT WAS NOT IMPACTED BY THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC SYNCHRON ® CHEMISTRY ANALYZER DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER FOR CLINICAL USE JJE BECKMAN COULTER INC. DXC 800 PRO N/A

Patients

Seq Age Sex Outcome Treatment
1