UNICEL® DXC SYNCHRON ® CHEMISTRY ANALYZER
Report
- Report Number
- 2050012-2011-04702
- Event Type
- Malfunction
- Date Received
- August 25, 2011
- Date of Event
- July 25, 2011
- Report Date
- July 25, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
SAMPLE INFORMATION AS NOT PROVIDED. THE CUSTOMER REPORTED TO CUSTOMER TECHNICAL SUPPORT (CTS) THAT THEY OBSERVED THE BUNM ELECTRODE HAD THE WRONG QUAD RING ATTACHED (DESCRIBED AS THINNER AND SMALLER BY CUSTOMER) AND IT WAS LEAKING. THE CUSTOMER REPLACED THE ELECTRODE WITH ONE THAT HAD THE CORRECT QUAD RING. CTS SENT THE CUSTOMER A REPLACEMENT ELECTRODE. THE CUSTOMER DID NOT GENERATED SERVICE REQUEST.
THE CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) TO REPORT ERRONEOUS BUNM RESULTS GENERATED BY THE UNICEL DXC 800 PRO SYNCHRON CHEMISTRY ANALYZER. NO ERRONEOUS RESULTS WERE REPORTED OUT OF THE LABORATORY. THE ERRONEOUS SAMPLES WERE REPEATED ON ANOTHER INSTRUMENT; HOWEVER, THE CUSTOMER WAS UNABLE TO PROVIDE DATA. PATIENT TREATMENT WAS NOT IMPACTED BY THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC SYNCHRON ® CHEMISTRY ANALYZER | DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER FOR CLINICAL USE | JJE | BECKMAN COULTER INC. | DXC 800 PRO | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |